5.2 Diagnostic LED
Current consumption values:
I
Q
= 209
uA
Quiescent current
(5.1)
I
W
= 10
mA
Wake current
(5.2)
I
M
= 55
mA
Measure current
(5.3)
Current consumption timing:
t
Q
= 1
−
(
t
W
+
t
M
)
(5.4)
t
W
=
3
s
60
s
(5.5)
t
M
=
5
s
n
×
60
s
(5.6)
n
= number of minutes between samples
(5.7)
Current consumption formula:
I
Avg
=
t
Q
×
I
Q
+
t
W
×
I
W
+
t
M
×
I
M
(5.8)
Figure 5.3: Current profile
As an example for determining battery life of a 3AH battery with sampling every 5 minutes:
T
Batt
=
AH
Rating
I
Average
(5.9)
=
AH
Rating
(
I
Q
×
t
Q
+
t
W
×
I
W
+
t
M
×
I
M
)
(5.10)
=
3
AH
(209
uA
×
0
.
9 ˙3 +
3
s
60
s
×
10
mA
+
5
s
5
×
60
s
×
55
mA
)
(5.11)
= 1861 hrs, or 78 days, or 2.5 months
(5.12)
Note that measuring is still performed for ”Threshold” settings, even when it is too bright to record a reading. Measuring
is performed to determine this ”measure” state, and
I
M easure
still applies.
5.2 Diagnostic LED
When the unit is connected to a battery, the red LED near the USB connector will illuminate in the sequence shown in
Figure 5.4 (firmware feature version 26 onwards):
Unihedron SQM-LU-DL Operator’s Manual -
20220119
21