8.5 Datalogger commands
Table 8.29 –
continued from previous page
Mode
Description
LM5x
Logging every 15 minutes on the 1/4 hour, and powering down between recordings.
LM6x
Logging every 30 minutes on the 1/2 hour, and powering down between recordings.
LM7x
Logging every hour on the hour, and powering down between recordings.
The response of setting the mode is simply an acknowledgement of the trigger mode. Also, reading the mode can be
done with “
Lmx
”, both will produce the response shown in table 8.30:
Table 8.30: Response of setting or getting the log trigger mode
Column
Example value
Description
0-2
LM,
Confirmation of command.
3
0
0 = Off
1 = Seconds
2 = Minutes
3 = Every 5 minutes on the 1/12th hour
4 = Every 10 minutes on the 1/6th hour
5 = Every 15 minutes on the 1/4 hour
6 = Every 30 minutes on the 1/2 hour
7 = Every hour on the hour
4-5
Carriage return (0x0d), Line feed (0x0a).
An example response is:
LM,0
012345
8.5.11 Logging Interval setting response
Either making the request “LIx” or any request to set the logging interval report setting produces the response shown in
table 8.31:
Table 8.31: Response of viewing or setting logging interval reporting parameters
Column
Example value
Description
0-1
LI
Logging interval settings from EEPROM and RAM are being re-
turned.
3-13
0000000360s
Logging interval period in
seconds
that was set into EEPROM
15-25
0000000005m
Logging interval period in
minutes
that was set into EEPROM
27-37
0000000121s
Current interval period in
seconds
from RAM
39-49
0000000004m
Current interval period in
minutes
from RAM
51-62
00000017.60m
Threshold value that was set into RAM
63-64
Carriage return (0x0d), Line feed (0x0a).
An example response is:
Unihedron SQM-LU-DL-V Operator’s Manual -
20211013
79