8 Commands and responses
Table 8.14: Response for manually setting of light calibration temperature
Column
Example value
Description
0
z
Calibration response is being returned.
2
6
Manual Set Light Calibration Offset
4-9
019.0C
Value that was set into EEPROM
10-11
Carriage return (0x0d), Line feed (0x0a).
An example of the response is:
z,6,019.0C
0123456789 1
8.3.7 Manually set dark calibration time period
Calibration is done at the factory, however, in the case where calibration values must be restored or set to something else,
this command allows a new calibration value to be placed into the meter.
The “Dark calibration time period” is the amount of time that has elapsed for the light sensor to make one cycle while
in complete darkness. The meter sets a time limit of 300 seconds on this value.
Executing the command “
zcal7#######.###x
” manually sets the light calibration offset to the value specified in
“
#######.###
”. The units are in seconds.
The format of the response is shown in table: 8.15:
Table 8.15: Response of manually setting dark calibration time period
Column
Example value
Description
0
z
Calibration response is being returned.
2
7
Manual Set Dark Calibration time period.
4-15
0000300.000s
Value that was set into EEPROM
16-17
Carriage return (0x0d), Line feed (0x0a).
An example of the response is:
z,7,00000300.00s
0123456789 1234567
8.3.8 Manually set dark calibration temperature
Calibration is done at the factory, however, in the case where calibration values must be restored or set to something else,
this command allows a new calibration value to be placed into the meter.
The “Dark calibration temperature” is the temperature of the meter when the meter was calibrated for its “Dark
calibration time period”.
Executing the command “
zcal8########.##x
” manually sets the light calibration offset to the value specified in
“
########.##
”. The units are
º
C.
Note: The meter records the temperature in a raw value with different resolution, so the reply back may not be exactly
the same as the value sent.
The format of the response is shown in table: 8.16:
70
Unihedron SQM-LU-DL-V Operator’s Manual -
20211013
Summary of Contents for SQM-LU-DL-V
Page 2: ......