10
T
08 9351 7500
E
W
www.unicarehealth.com.au
Maintenance Guide Check List
•
General visual inspection of lifter.
•
Check castors are tightly attached to legs.
•
Check castors run & swivel freely, remove any hair & fl uff from axle.
•
Check castor brakes operate effi ciently.
•
Over haul castors & pack with grease.
•
Inspect surfaces for scratches & chips
•
Clean surfacers with a neutral detergent, remove any adhesive tape with orange or eucalyptus oil.
•
Examine all moving parts and fasteners for wear and fatigue, lubricate and /or adjust as required.
•
NB leg pivot bushes incorporate a Tefl on coated pivot bush which requires minimal lubrication .
•
Check tightness of bolt in leg bushes.
•
Check linkages are secure in leg adjuster ball joints.
•
Check all electrical components for correct operation.
•
Check battery charger operation.
•
Remove dirt & dust from actuator.
•
Wipe actuator shaft with oiled rag (DO NOT use WD40 or equivalent.)
•
Check mast to boom pivot point, remove bolts and lightly regrease bush and retighten, ensuring
free movement at pivot.
A MAJOR SERVICE SHOULD BE CARRIED OUT EVERY 12 MONTHS AS A MINIMUM REQUIREMENT.
IMPORTANT: An annual load test to the safe working load(swl) is required for
compliance with AS ISO 10535 − 2002