SET-UP
The EMV+ can be configured to suit most applications. Additional hoses, fittings and adapters may be
required for particular uses.
1.
For use with external O
2
: connect a green (white for some installations) high-pressure O
2
hose to
the OXYGEN IN fitting and a 55 psig external source.
NOTE:
Use only with medical-grade (USP) oxygen. When using with an oxygen cylinder, the cylinder must
be secured.
2.
Connect the disposable ventilator circuit to the GAS OUTPUT, TRANSDUCER, and EXHALATION
VALVE connectors on EMV+ Connector Panel. Follow the directions included with the disposable
ventilator circuit.
3.
In a high-dust or biological environment, a bacterial/viral filter should be attached to the Fresh
Gas/Emergency Air Intake to prevent entrainment of particulate or biological matter. (See
section on Hazardous Environment Filters.)
4.
In toxic biological or chemical environments the user can attach a chemical/biological filter to the
Fresh Gas/Emergency Air Intake. The threaded interface accommodates chemical/biological
filters with an Rd 40 x 1/7 interface (see BS EN 148-1 1999 Respiratory protective devices –
threads for face pieces). (See section on Hazardous Environment Filters.)
5.
For use with AC power: connect the AC/DC Power Supply (supplied) to the External Power Input
(see Figure 2).
NOTE:
The EMV+ can operate from internal battery or from external AC or DC power sources. See
section entitled Operating Power Selection & Stopping for details.
WARNING!
Always assure that there is an alternate means of providing ventilation. A bag-valve
resuscitator and an appropriate mask for the patient being ventilated should be immediately available.
WARNING!
This unit’s USB connection (unlabelled) does not provide any signal output or input to the
Operator. It is for the use of Impact Instrumentation trained personnel only. The Operator should not
connect anything to the USB connection.
HIGH PRESSURE
OXYGEN INPUT
GAS
OUTPUT
EXTERNAL POWER
INPUT
USB
CONNECTOR
TRANSDUCER
(PATIENT
AIRWAY
PRESSURE)
PULSE
OXIMETER
CONNECTOR
EXHALATION
VALVE
FRESH
GAS/EMERGENCY
AIR INTAKE