Chapter 5 Troubleshooting 

 

Possible  faults  in  use  of  UT1000X  and  troubleshooting  methods  are  listed  below.  Please  handle  fault  as  the 

corresponding steps. If it cannot be handled, please contact with the distributor or local office and provide the 

model information (press Utility 

System

 About to check).

 

 

5.1 

No Display on Screen(Blank Screen) 

 

If the waveform generator is blank screen when press the power switch on the front panel. 

1) 

  Inspect whether power source is connected well.

 

2) 

 

Inspect whether power button is pressed.

 

3) 

 

Restart the instrument

4) 

  If  the  instrument  still  can’t  work,  please  contact  with  the  distributor  or  local  office  for  product 

maintenance service.

 

 

5.2 

No Waveform Output 

In correct setting but the instrument has no waveform output display.

 

1) 

  Inspect whether BNC cable and the output terminal is connected well

 

2) 

 

Inspect whether CH1, CH2 button is turned on

3) 

  If the instrument still can’t work, please contact with the distributor or local office for product 

 

maintenance service.

 

Chapter 6 Service and support   

6.1

 

 

Maintenance and Cleaning

 

(1)

 

General Maintenance 

Keep the instrument away from the direct sunlight. 

    Caution     

Keep sprays, liquids and solvents away from the instrument or probe to avoid damaging the instrument 
or probe. 

 

(2)    Cleaning 

       

Check  the  instrument  frequently  according  to  the  operating  condition.  Follow  these  steps  to  clean  the 

external surface of the instrument: 

a. Please use a soft cloth to wipe the dust outside the instrument. 

b. When cleaning the LCD screen, please pay attention and protect the transparent LCD screen.   

Summary of Contents for UTG1000X Series

Page 1: ...User s Manual UTG1000X Series Function Arbitrary Waveform Generator...

Page 2: ...and its subsidiaries or suppliers which are protected by national copyright laws and international treaty provisions Information in this manual supersedes all previously published versions UNI T is t...

Page 3: ...d tear improper use and improper or lack of maintenance UNI T under the provisions of this warranty has no obligation to provide the following services a Any repair damage caused by the installation r...

Page 4: ...llow the common safety procedures Safety Precautions Warning Please follow the following guidelines to avoid possible electric shock and risk to personal safety Users must follow the following convent...

Page 5: ...be met before you proceed to operation Note It indicates potential problems which may cause failure of this device if you fail to follow a certain procedure or condition If the Note sign is present al...

Page 6: ...s a registered trademark of UK Certification ETL indicates a registered trademark of Intertek It conform to UL STD 61010 1 and 61010 2 030 CSA STD C22 2 No 61010 1 and 61010 2 030 Waste This product c...

Page 7: ...lectricity The protection grade of this device is 4 kV for contact discharge and 8 kV for air discharge Measurement accessories Measurement accessories are of lower class which are definitely not appl...

Page 8: ...k of this device Do not allow any external objects to enter this device via ventilation holes Please ensure adequate ventilation and leave a gap of at least 15 cm on both sides front and back of this...

Page 9: ...cification of input AC power Voltage Range Frequency 100 240VAC fluctuant 10 50 60Hz 100 120VAC fluctuant 10 400Hz Please use the attached power lead to connect to the power port Connecting to service...

Page 10: ...nction arbitrary waveform generator supports communication with the computer via USB interface User can use SCPI via USB interface and combined with programming language or NI VISA to remote control t...

Page 11: ...y 2 1 Main Features Frequency output of 40MHz 20MHz 1 Hz resolution of full frequency band DDS function it has 200MSa s sampling rate and 16 bit vertical resolution Square wave with the maximum freque...

Page 12: ...the current input 4 Multifunction Knob Arrow Key Multifunction knob is used for changing number rotate clockwise to increase number or as the arrow key press the knob to select the function or to con...

Page 13: ...d number of cycles When the pulse string mode is gated the gating signal is input through the external digital modulation interface When using the frequency meter function the signal compatible with T...

Page 14: ...sed by the company 3 Safety Lock Safety lock sold separately can be used for the instrument stay at fixed position 4 AC Power Input Interface AC power specification of UTG1000X function arbitrary wave...

Page 15: ...ys and multifunction knob If the bottom color of the current character is the color of the current channel it is white when the system is in setting it means that this character enters the editing sta...

Page 16: ...oltage of the waveform is 0V sine wave in default connect with 50 port The specific steps to change DC offset voltage to 150mV Press Wave Sine Offset key in turn use numerical keyboard to input 150 an...

Page 17: ...al keyboard to input 1 and then select the unit of the parameter to V 3 1 9 Power Output The full bandwidth of built in power pre amplifier can reach to 100 kHz the maximum output power 4W output slew...

Page 18: ...uency Meter This function arbitrary waveform generator can measure the frequency and duty cycle of compatible TTL level signals The range of measurement frequency is 100mHz 200MHz When use the frequen...

Page 19: ...ol ArbMng Interna OK Select Export ALL the arbitrary wave in the current folder is exported to the external storage device External arbitrary wave Import Select the Utility Tool ArbMng External OK sel...

Page 20: ...to the system selection menu System there are two pages you need to press Next key to turn page 1 Start Phase Select PhaseSync to Independent or Sync Independent The output phase of CH1 and CH2 output...

Page 21: ...e computer via USB b Press and hold Utility knob to turn on the power supply of signal source and then release the button c Use write tool to write the firmware to the signal source and then restart t...

Page 22: ...r Carrier Wave The frequency range of the carrier wave can be set differently The default frequency of the carrier wave is 1 kHz The frequency of each carrier wave can refer to the table as follows Ca...

Page 23: ...s 100 Falling Ramp Wave Symmetry is 0 Arbitrary Wave The length of arbitrary wave limits at 4 kpts by the method of select point automatically when arbitrary wave is the modulation wave Noise Wave Whi...

Page 24: ...internal 200 Hz sine wave as a modulating signal another square wave signal with frequency 10 kHz amplitude 200 mVpp duty ratio 45 as the carrier wave and set modulation depth to 80 at last the settin...

Page 25: ...o input 200 and then select the unit of parameter to mVpp Press Duty soft key to set the duty cycle use numerical keyboard to input 45 and then select the unit of parameter to As shown in the followin...

Page 26: ...through an oscilloscope as shown in the following figure 4 1 2 Frequency Modulation FM In FM mode modulated wave is consist of the carrier wave and the modulation wave The frequency of the carrier wav...

Page 27: ...n source of this instrument comes from internal it is sine wave square wave rising ramp wave falling ramp wave arbitrary wave and noise wave The default is sine wave When FM mode is enabled the defaul...

Page 28: ...frequency deviation is 1 kHz It can be change by ModFreq soft key in FM mode Frequency deviation Carrier wave frequency if the frequency deviation is greater than the carrier wave frequency the instru...

Page 29: ...tion wave press ModFreq soft key and use numerical keyboard to input 2 and then select the unit of parameter to kHz 3 Set Wave and Parameter for Carrier Wave Signal Press Wave key to enter the carrier...

Page 30: ...it of parameter to mV 4 Set Frequency Deviation After the parameter setting of the carrier wave is completed press Mode Shape FM to enter FM mode setting Press FreqDev soft key and use numerical keybo...

Page 31: ...se Modulation PM In PM mode modulated wave is consist of the carrier wave and the modulation wave The phase of the carrier wave will changed with the amplitude of the modulation wave Press Mode Shape...

Page 32: ...e rising ramp wave falling ramp wave arbitrary wave and noise wave The default is sine wave When PM mode is enabled the default modulation wave is sine wave It can be change by multifunction rotary kn...

Page 33: ...is 180 It can be change by PhaseDev soft key and arrow keys in PM mode Comprehensive Example In PM mode set an internal 200 Hz sine wave as a modulating signal another sine wave signal with frequency...

Page 34: ...Hz Press Amp soft key to set the amplitude use numerical keyboard to input 100 and then select the unit of parameter to mVpp As shown in the following figure 4 Set Phase Deviation After the parameter...

Page 35: ...the PM through an oscilloscope as shown in the following figure 4 1 4 Amplitude Shift Keying ASK ASK is a digital amplitude modulation it expresses digital signals 0 and 1 by changing the amplitude of...

Page 36: ...er wave can be sine wave square wave ramp wave or arbitrary wave except DC The default is sine wave After ASK mode is selected press Wave key to enter the carrier wave interface Set Frequency for Carr...

Page 37: ...e logical level of external input is low it outputs the hopping frequency when the logical level of external input is high Set ASK Rate When ASK mode is enabled it can set ASK rate the range is 2 mHz...

Page 38: ...the default is sine wave Press Freq soft key to set the frequency use numerical keyboard to input 2 and then select the unit of parameter to kHz Press Amp soft key to set the amplitude use numerical...

Page 39: ...shown in the following figure 4 1 5 Frequency Shift Keying FSK In FSK mode it can set the switch rate between the frequency of carrier wave and hopping frequency for the instrument Select FSK Press Mo...

Page 40: ...ifunction rotary knob or press ModSrc External soft key 1 Internal Source When the modulation source is internal the internal modulation wave is a square sine with duty ratio 50 non adjustable The fre...

Page 41: ...arrier wave and hopping frequency When FSK mode is enabled it can set FSK rate the range is 2 mHz 100 kHz The default is 100 Hz It can be change by multifunction rotary knob and arrow keys or press so...

Page 42: ...d to input 200 and then select the unit of the parameter to Hz 4 Set Carrier Wave Signal Press Wave key to enter the carrier wave interface and select sine wave to be the carrier wave the default is s...

Page 43: ...CH1 key if the indicator light is illumined which means the channel output is enabled To view the shape of the FSK through an oscilloscope as shown in the following figure 4 1 6 Phase Shift Keying PS...

Page 44: ...he currently setting Select Carrier Wave The carrier wave can be sine wave square wave ramp wave or arbitrary wave except DC The default is sine wave After PSK mode is selected press Wave key to enter...

Page 45: ...al input is high Set PSK Rate When the modulation source is internal it can set the frequency of movement between the phase of carrier wave and modulation phase When PSK mode is enabled it can set PSK...

Page 46: ...rd to input 500 and then select the unit of the parameter to Hz 3 Set Phase Press soft key Phase and use numerical keyboard to input 180 and then select the unit of the parameter to 4 Set Carrier Wave...

Page 47: ...the unit of parameter to kHz Press Amp soft key to set the amplitude use numerical keyboard to input 1 and then select the unit of parameter to Vpp 5 Enabling Channel Output Press CH1 key if the indi...

Page 48: ...amplitude of the modulation wave The modulation mode of the two channelsare independent user can set the same or different modulationmode for thetwo channels Select PWM Press Mode Shape PWM to turn on...

Page 49: ...n be set to 0 49 99 the default is 20 It can be change by multifunction rotary knob and arrow keys or press DutyCycle key Notes Duty cycle deviation indicates the change in duty cycle of the modulated...

Page 50: ...lation Rate Press Rate soft key and use numerical keyboard to input 200 and then select the unit of the parameter to Hz 3 Set Duty Cycle Press DutyCycle soft key and use numerical keyboard to input 50...

Page 51: ...frequency use numerical keyboard to input 2 and then select the unit of parameter to kHz Press Amp soft key to set the amplitude use numerical keyboard to input 1 and then select the unit of paramete...

Page 52: ...all generate the frequency sweep output 4 2 1 Select Frequency Sweep 1 Enabling Frequency Sweep Mode Press Mode Sweep Linear to turn on frequency sweep mode the instrument will output the frequency sw...

Page 53: ...DS function arbitrary waveform generator sweeps from low frequency to high frequency When start frequency stop frequency DDS function arbitrary waveform generator sweeps from high frequency to low fre...

Page 54: ...use numerical keyboard to input number and select unit soft key to complete the setting 4 2 5 Comprehensive Example In frequency sweep mode set an internal square wave with amplitude 1 Vpp duty cycle...

Page 55: ...unit of parameter to the default duty cycle is 50 3 Set Start Stop Frequency Sweep Time Press Mode Sweep Linear to enter linear sweep Press Start or Stop soft key use numerical keyboard to input 1 an...

Page 56: ...internal and external trigger to control output the pulse string Pulse string has three types N cycle gating and infinite Sine wave square wave ramp wave pulse wave arbitrary wave except DC or noise...

Page 57: ...h the above steps press Wave key to enter the carrier wave interface 3 Set Wave Frequency In N cycle and gating mode wave frequency is defines the signal frequency during pulse string In N cycle mode...

Page 58: ...se String Type UTG1000X can output three types of pulse train N cycle gating and infinite The default type is N cycle 1 N Cycle Mode Press NCycle to enter N cycle mode the waveform generator will outp...

Page 59: ...rity to complete the setting as shown in the following figure 3 Infinite Mode Press Infinite to enter infinite mode burst period pulse string period and cycle number options in the parameter list will...

Page 60: ...knob and arrow keysor press soft key Burst to complete the setting Notes Burst period of pulse string waveform period cycle number number of pulse string At here waveform period is the reciprocal of...

Page 61: ...tring of N cycle and infinite type 4 3 7 Trigger Edge The external digital modulation terminal FSK CNT connector can be designated edge as input or output The rising edge and falling edge can be set r...

Page 62: ...rrow keys Remarks If the setting interface displays Freq this means it can only change the frequency parameter it can t convert to period If it displays Period then 2ms period is corresponding to 500...

Page 63: ...t numerical value and then select the unit 4 Enabling Channel Output Press CH1 key on the front panel to quickly turn on CH1 output or press Utility key to pop out the label and then press soft key CH...

Page 64: ...e Press Mode Arbitrary to turn on arbitrary wave function the instrument will output the arbitrary waveform according to the current setting 4 4 2 Select Arbitrary Wave User can select the built in ar...

Page 65: ...t maintenance service 5 2No Waveform Output In correct setting but the instrument has no waveform output display 1 Inspect whether BNC cable and the output terminal is connected well 2 Inspect whether...

Page 66: ...etailed provisions of the warranty To arrange for repair or acquire warranty form please contact the nearest UNI T sales and repair department In addition to permit provided by this summary or other a...

Page 67: ...Upper Limit of Amplitude 10 V Lower Limit of Amplitude 10 V Fundamental Wave Frequency 1 kHz Amplitude 100 mVpp DC Deviation 0 mV Start Phase 0 Duty Cycle of Square Wave 50 Symmetry of Ramp Wave 50 D...

Page 68: ...ate 100 Hz FSK Modulation Source Internal FSK Rate 100 Hz Hopping Frequency 10 kHz PSK Modulation Source Internal PSK Rate 100 Hz Phase 180 PWM Modulation Wave Sine wave PWM Rate 100 Hz Duty Cycle 20...

Page 69: ...sian_monopulse Gaussian monopulse GaussPulse Gaussian pulse NegRamp Descent slope NPulse N Pulse signal PPulse P Pulse signal SineTra Tra sine signa SineVer Ver sine signal StairUD Ladder ladder Stair...

Page 70: ...ction Besselk Besselk function Bessely Class II Bessel function Cauchy Cauchy distribution Cubic Cubics function Dirichlet Dirichlet function Erf Error function Erfc Complementary error function ErfcI...

Page 71: ...automobile starting oscillation ISO16750 2 Starting1 Automobile starting voltage waveform 1 ISO16750 2 Starting2 Automobile starting voltage waveform 2 ISO16750 2 Starting3 Automobile starting voltage...

Page 72: ...ion Tan Tangent function TanH Hyperbolic tangent AntiTrigonome 16 types ACosH Arc hyperbolic cosine function ACotCon Arc hyperbolic cosine function ACotPro Convex arc cotangent function ACotHCon Conca...

Page 73: ...The minimum of four Blackman Harris window ParzenWin Parzen window TaylorWin Taylaor window Triang Quarter window Fejer window TukeyWin Tukey window Complex Wavelets 7 types Complex Frequency B spline...

Page 74: ...User s Manual UTG1000X Series Instruments uni trend com 74 74 KS_1_1 MAF_1_1 MAF_1_2 MAF_5_3 MAP_1_1 MAP_1_2 MC_3 Mexican hat Mexican hat wavelet O2PROPA1 O2PROPA2 O2SNAP STAR02_1 TPS_1_1 TPS_1_2...

Reviews: