Model UT71A/B/C/D/E: OPERATING MANUAL
K. 4~20 mA loop current (UT71B/C/D/E)
Range
Resolution
Accuracy
Overload Protection
(4~20mA)%
0.5A, 1000V, fast type fuse, 6.3x32mm
(1%+50)
0.01%
Remarks:
< 4mA, the primary display shows LO
4mA, the primary display shows 0% .... ; 20mA, the primary display shows 100%
> 20mA, the primary display shows HI
** END **
When the readings obtained is:
L. Power Measurement (UT71E only)
Range
Resolution Accuracy Current Overload Protection
2500W
1000V
0 . 1 W
10A, 1000V, fast type fuse,
10.3x38mm
Voltage Overload Protection
POWER
VA POWER
Cos POWER
2500W
1W
(2%+10)
(1%+10)
(2%+10)
Remarks:
Power factor input range: 0.000~1.000
Voltage input impedance: around 10M .
Voltage input range: AC50~250V
0 ~ 1
0.001W