Accuracy: (a% r b digits), guarantee for 1 year.
Operating temperature: 23
o
C 5
o
C.
Relative humidity: <75%.
Temperature coefficient: 0.1 x (specified accuracy) / 1
o
C.
When measuring 20
Ω
and 200
Ω
range, the test clips
can add 0.1 to 0.3
Ω
error to resistance. To obtain precise
readings in these low-resistance measurement, that is
the range 20
Ω
and 200
Ω,
short circuit the input terminals
beforehand and record the reading obtained (called this
reading as X). (X) is the additional resistance from the
test lead.
Then use the equation:
measured resistance value (Y) – (X) = precision readings
of resistance
l
19
Model UT601: OPERATING MANUAL
Accuracy Specifications
A. Resistance Test
Range Resolution
Accuracy
Overload
Protection
20
Ω
200
Ω
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
200M
Ω
2000M
Ω
0.01
Ω
0.1
Ω
0.001k
Ω
0.01k
Ω
0.1k
Ω
0.001M
Ω
0.01M
Ω
0.1M
Ω
1M
Ω
(1%+5)
(0.8%+3)
(0.8%+1)
(1%+2)
[5%(rdg-10)+10]
Reference only
250V rms
Remarks: