background image

P/N:110401104398X

6. Rotary knob set to 500V/1000V/2500V:

    to select test voltage for insulation resistance measurement.

If low battery indicator shows on upper left corner of LCD after the 

meter is turned on, it means battery is almost used up and need to

be replaced.

(See Figure 1)

Figure 1

(1)  Set the rotary knob to ACV.

(2)  Insert the red test lead into "V" jack and the black test lead into 

      "G" jack.

Disconnect test leads and tested circuits and remove test leads 

away from input jacks after completing the measurement.

Please take extreme caution in order to avoid electric shock when 

measuring high voltage.

(See Figure 2)

With the instrument connected to tested circuit as described above,

lead

lead

test probe with the circuit 

under test.

Figure 2

(See Figure 3)

Figure 3

 ● 

Turn the knob to select test voltage 500V/1000V/2500V.

Do not short-circuit the test leads under high-voltage output status or 

make insulation measurement after high voltage has already been 

output.

lead

lead

With the knob already set to test voltage 500V/1000V/2500V, then 

press TEST button,the instrument will be self-locked to measure 

continuously. The test voltage will be output and TEST button will 

light up. With the measurement finished, press TEST button to 

unlock and stop the measurement.

When red and black clips are connected to the circuit under test, 

high voltage is output from LINE jack and the current from EARTH 

jack.

Dielectric absorption ratio measurement

Under insulation measurement mode, press PI/DAR button two 

times to select DAR parameter, the screen displays Time 1(30s)/

Time2(1min), then press TEST to start DAR measurement. Press 

TEST three times to select another DAR,  the screen shows Time 

1(15s)/Time2(1min), press to begin another DAR measurement.

Under insulation measurement mode, press PI/DAR button once 

to select PI parameter, the screen displays Time 1(1min)/

Time2(10mins), then press TEST to start PI measurement.

replacing

batteries

installing

set the knob to

the housing

To avoid damage to the instrument, do not submerge it into 

the water.

Clean the instrument surface with soft cloth or sponge 

dampened with clean water.

When it is necessary to verify or repair instrument, please

deliver the instrument to qualified professional serviceman

or designated repairing department. 

M

R

*END*