background image

      adjustable; long press “ME

NU”

 to enter function selection menu, and then             

      short press it to select SECREC interface, while the auto memory interval is 

      displayed on the main display; press “RA

N

GE/+” or “M

AX/

MI

N

/-” key to adjust 

      the auto memory interval;  press “ME

NU”

 key to enter 

nex

t function setting or 

      press “H

O

LD” to 

exi

t function setting. And luminometer will save the updated 

      settings.

  

  Long press “REC/ERASE” to save data automatically, and “REC” sign 

      on secondary display 

flashes

, and the luminometer will save the 

      measured data according to the current auto saving interval setting; if the 

      storage space of luminometer is fully taken up, the system will 

exi

t auto 

      memory; and “FULL” sign will be displayed on LCD; press “REC/ERASE” key 

      to 

exi

t auto data memory function.

  

  Manual data memory: in the normal measurement state, press “REC/ERASE” 

      once to save one measured data, and “REC” sign displayed on secondary 

display 

flashes

 once.

  

  When 2044 pieces of data is fully recorded, “FULL” sign will be displayed on 

       LCD, and no data can be saved at this moment. 

  

  Under power-off state, press “REC/ERASE” and power on meanwhile, when 

      “CLR” sign is displayed on LCD, all saved data will be cleared.

3. View saved data:

  

  Short press “ME

NU”

 to enter view recorded data function; if there is no 

      data record in the luminometer, “

----

” sign will be displayed on both LCD 

      main display and secondary display, and the luminometer will return to normal 

      measurement state in about 0.5 seconds; If there a certain data records in 

      the luminometer, the total number of current data records will be displayed on 

      LCD secondary display, and the value of the last piece of data will be 

      displayed on the main display.

  

  After entering option of view recorded data, press 

“UN

IT” to view the 1st 

      piece of recorded data; press “RA

N

GE/+” and “M

AX/

MI

N

/-” keys to view data 

      records forward or backward; 

each press on HOLD button to increment by 

     100 sets, the meter will return to the first record when the maximum number 

      has been reached.

  

  The luminometer can store up to 2044 records.

4.   Restore to factory defaults:

  

  Long press “ME

NU”

 key to enter function setting, and then short press this 

      key to select 

"DEF?" option, 

 and “DEF

?”

 sign displays on the LCD and 

      

flashes

, at this moment you can press “H

O

LD” to restore to factory defaults; 

      press “ME

NU”

 or “REC/ERASE” 

to exit.

 

  

  After restoring to factory defaults, the defaulting state of the system is: USB.0 

      (no USB transmission); 

APO

1 (auto power off); 60s (auto record interval is 

      set up as 60s); clearing all data records.

5.USB data transmission function: (only for UT382, See Figure 4)

  

  Long press “ME

NU”

 to enter USB 

      transmission setting, and “USB.0” or

      “USB.1” will be displayed on LCD;

      USB.0 means that data can’t be

      transmitted; “USB.1” means that data

      can be transmitted; press “RA

N

GE/+”

       or  “M

AX/

MI

N

/-” key to switch between

       “USB.0” and “USB.1”.     
  

  The meter is set to "USB.0" mode 

      automatically every time it is powered on.
  

  The luminometer (UT382) communicates 

      with PC using USB wire and your PC must 

      have a USB port. See Figure 4 for the 

      connection diagram.
  

  When connecting UT382 to the computer, 

      you cannot operate any functional buttons 

      during transmission. 

Figure 4

6. Illuminance measurement

  

  In process of measuring illuminance, the sensitive mirror must be 

      perpendicular to the measured illuminating source to ensure accuracy of 

      measured data..
  

  If distance between measured illuminating source and illuminometer, 

      measuring position, angle or environment  is different, then the measured 

      data will be different; because the intensity of light varies with change of 

      distance, position, angle and environment. 
  

  Make sure there is no scratch or dirt on sensitive mirror of luminometer, or 

      the measured data won’t be accurate.
  

  When the illuminance 

exceeds

 the measuring range of the luminometer, “

O

L” 

      sign will be displayed.
  

  After completing measuring, cover the sensitive mirror with cap to avoid 

      scratch on sensitive mirror and keep it clean.

VI Specifications

1. General 

specifica

tion

  

  Low battery indication:

  

  Type of sensor: digital visible light sensor

  

  Sampling rate: 100 times/s

  

  Power supply: one 1604A 6F22 9V battery

  

  Battery life: 200 hours for typical battery (alkaline battery)

  

  Dimensions: 

195mmx

 

45mmx

 26m

  

  Weight: about 185g (including battery)

2. Environment limit

  

  Indoor use

  

  Safety standards: E

N61326

: 2006; 

                                   E

N55022

: 1998+A1+A2;

                                   E

N55024

:  1998+A1+A2

  

  Pollution degree 2

  

●  O

perating temperature and humidity

:

             0℃-30℃ 

 

(no more than 80%RH);

 

          

30℃-40℃

 

(no more than 75%RH);

 

          

40℃-50℃

 

(no more than 45%RH)

  

  Storage temperature and humidity

: -20℃-+60℃

 

(no more than 80%RH)

3. Electrical 

specifica

tion

  

  Accuracy tolerance: ± (a%rb digits), calibrate once each year 

  

  Environment temperature: 

23℃±5℃

  

  Environment humidity

: ≤

80% RH

  

  Temperature coef

ficien

t: 0.1× (accuracy tolerance)/

Illuminance measurement

 Function                Range       Resolution   

Accuracy Tolerance: ± ( a %rb digits)

Illuminance 

measuring 

(LU

X

)

Illuminance 

measuring 

(FC)

The accuracy tolerance of FC can 

be 

verified

 by unit conversion:

FC=10.

76lux

, if accuracy tolerance 

verifica

tion is needed, it can be 

achieved by unit conversion.

20Lu

x

200Lu

x

2000Lu

x

20000Lu

x

2FC
20FC
200FC
2000FC

0.01Lu

x

0.1Lu

x

1Lu

x

10Lu

x

0.001FC
0.01FC
0.1FC
1FC

± (3% + 20)
± (3% + 8)
± (3% + 8)
± (3% + 8)

VII Maintenance

         Warning

During replacing battery or measuring, keep the sensitive mirror from scratch or 

dirt to avoid damaging the mirror or 

influencing

 the measuring precision; and do 

not charge the replaced battery to avoid 

explosion

 and safety accident!

1. Battery installation and replacement

When “      ” is displayed on the luminometer, replace the battery immediately.

Replace the battery by following procedures:

  

  Power off

  

  Loosen the screw of battery cover and take it off. 

  

  Replace with a new 6LF22 9V 1604A battery, please use battery with the 

      same model, and do not use unauthorized battery.

  

  Pay attention to the anode “+” and the cathode “-”. After completing 

      installation, install the cover and tighten the screw.
2. General maintenance

  

  When the surface of luminometer is dirty, clean with wet cloth and neutral 

      detergent, grinding miller and solvent are forbidden.

  

  When the luminometer is damaged and needs to be repaired, please send it   

      to the designated maintenance center of our company for repair by 

      professional service man, do not repair it without authorization.

  

  Especially in the process of using, keep the sensitive mirror of luminometer 

      away from scratch or dirt to avoid damaging the mirror or 

influencing

 the 

      measuring precision.

  

  Remove the battery if it won’t be used for a long time.

  

  Store the luminometer in a place free of moist, high temperature and strong 

      magnetic 

field

.

S

pecifica

tions and other information shown on this instruction manual are subject 

to change without notice

  

  LCD: 3 1/2-bit display, 2000-count.

  

●  O

ver range display: “

O

L”

P/N:110401104377X  DATE:2018.06.26  REV.5