I Overview

UT380 series luminometer has two models: UT381 and UT382. UT380 series 

luminometer is a kind of digital meter applying high-precision digital visible light 

sensor and 8-bit microprocessor processing data. 
Illuminance: Illuminance means the luminous 

flux

 received on each unit area 

of illuminated object, with adopted unit of foot candle (12 inches make 1foot) 

in Britain and USA and meter candle in Europe. 1 foot candle means the 

illuminance received on the surface from the direct 1 candle light source with 1 

foot away from the illuminated surface, abbreviated as FC. In the same way, 1 

meter candle means the illuminance received on the surface from the direct 1 

candle light source with 1 meter away from the illuminated surface, abbreviated 

as 

Lux

. UT380 series can test illuminance within 0

20,

000Lux

, and UT382 can 

be connected with PC to realize real-time data storage and analysis, and the data 

recorded in luminometer can be transmitted to PC to analyze, print and record, etc. 

II Unpacking Inspection

Unpack and check, if there is any damage or loss, contact with the nearest local 

sales service agency 
  1. Mainframe

1 pc

  2. Instruction manual

1 pc

  3. USB testing line (only for UT382) 

1 pc

  4. Software disc(only for UT382)

1 pc

  5. 9V battery

1 pc

  6. Sensitive mirror cap

1 pc

III Safety Instruction

       Warning
Actions and conditions that may pose hazard to the user or may cause 

damage to the Luminometer. 

please use the luminometer as specified in the manual, otherwise it may cause 

damage to the luminometer or personal injury. In order to avoid luminometer 

damage and personal injury, please operate according to following instructions.

  

  Check if luminometer shell is broken or any part is lost before using. Do not 

       use the meter if it is damaged.

  

  When battery indicating sign “         ” appears, please replace battery as soon  

      as possible to avoid measured data error.

  

  When the luminometer works abnormally, do not use it, and please send it to 

      the designated maintenance center of our company for repair by professional 

      personnel.

  

  Please do not use it in the place adjacent to 

explosive

 gas, steam and dust.

  

  Do not dismantle the shell of luminometer without authorization to avoid 

      luminometer damage.

  

  Do not charge the battery to avoid battery 

explosion

 and personal injury. 

      Please pay attention to the “+” and “-” of the battery during installing.

  

  Especially keep the sensitive mirror clean and away from scratch, cover it 

      with cap after completing measurement.

1     Sensitive mirror
2     Mainframe
3     Display screen
4     Keyboard

IV Meter Structure

1. Meter structure (Figure 1,Table 1)

Figure 1

Table 1

2. Display symbols (Figure 2, Table 2)

Figure 2

Auto range
Minimum value display

Maximum

 value display

Data hold
Main display
Full storage display
Auto power off
USB communication (UT382 only)
Low battery Indication
record number
Secondary display
Data record storage
Indicator for "reading ×10"
Illuminance unit 

(Lux

)

Illuminance unit (Britain and USA)

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

MIN

Auto

MAX

HOLD

FULL

USB

LoadNo

REC

Table 2

3.Keys description (Figure 3 )

Figure 3

1

2

3

4

5

6

7

1.   Power on/off key: long press to power on, shot press to power off.
2.   Unit selection key: under measuring state, press this key to switch the unit of 

      reading between 

Lux

 and FC

3.   

Maximum and minimum values

 selection key: under measuring state, press 

      the key to switch 

from normal measurement→ MAX→ MIN modes.

      When “M

AX

” or “MI

N”

 sign is displayed on LCD, the luminometer just displays 

      maximum or minimum reading.

 

4.   Auto range and Manual range selection key: in auto range state, press the key to 

      enter manual range state; press once again to enter low range and high range  

      one by one; long press the key to return to auto range state.

5.   Data memory and delete key: press this key to save measured data automatically 

      or manually, the luminometer can save up to 2044 sets. press this key and power 

      on the meter at the same time, you can delete the  recorded data.

6.   Menu key: long press this key to enter the function setting mode; short press to 

      access USB→APO→ SEC→ CODE→ DEF?→ Normal measurement;

 

      press H

O

LD key to 

exi

t menu mode.

7.   Data hold key: press this key to lock the measured data display on LCD, and 

      display “H

O

LD” sign at the same time; press this key again to 

exi

t hold function.

V  Setting and Operation

 

1. Auto power off: power off in about 10 minutes automatically

  

  Auto power off or not selection setting function: long press “ME

NU”

 to enter 

      function selection menu, and then short press it to select “

APO.

0” or “

APO.

1”  

      interface, and then press “RA

N

GE/+” or “M

AX/

MI

N

/-” key to switch between 

      “

APO.

0” and “

APO.

1”; press “ME

NU”

 key to enter 

nex

t function setting or 

      press “H

O

LD” key to 

exi

t function setting, and luminometer will record the 

      updated settings.
  

  

When "APO.0" is selected

, auto power off function is closed, namely, the 

      meter won’t be power off automatically; 

if "APO.1" is applied, au

to power

      off function is selected, and the auto power off sign is displayed on the 

      screen at the same time. After power off, the luminometer will record the 

      settings which will be restored after it is power on, it is unnecessary to set 

      up again at each time.

2. Data memory and delete: auto and manual data memory

  

  The interval of auto data memory is within 0.5-255 seconds, which is 