E. Continuity Measurement

 Range   Resolution       Remark

                                  Open Circuit Voltage around 

                                  -1.2V;Buzzer beeps at resistance 

                                  ≤10Ω; Buzzer not sound when 

                                  resistance > 70Ω;

F. Diodes Measurement 

 Range 

Resolution         Ove

rload Protection 

                        1mV               300Vr

ms 

Remark: Open Circuit Voltage around 3V, Displays 

approximate forward voltage drop.

G.  Capacitance

Range    Reso

lution   Accuracy    Overload Protection

 3nF          0.001nF    

           

 30nF 

0.01nF 

 

 

 300nF 

0.1nF 

 

 

 3μF 

1nF 

 

 

 30μF 

10nF 

 

 

 300μF 

100nF 

±(5%+5)   

 3mF 

 

reference 

Remarks: 

1. Under auto scan mode the max range is 300μF.

2. There is a residual reading when the circuit is open. 

To measure a small value of capacitance, subtract it 

to ensure accuracy.

Maintenance 

(see figure 4)

This section provides basic maintenance information 

and battery replacement instruction.

 

              

                                  Figure 4

         Warning

Do not attempt to repair or service your Meter unless 

you  are  qualified  to  do  so  and  have  the  relevant 

calibration, performance test, and service information.  

To avoid electrical shock or damage to the Meter, do 

not get water inside the case.

A. General Service

●   Periodically wipe the case with a damp cloth and 

     mild detergent.  Do not use abrasives or solvents.

●  To clean the terminals with cotton bar with 

    detergent, as dirt or moisture in the terminals can 

     affect readings.

●   Turn the Meter off when it is not use and take out 

     the battery when not using for a long time.

●   Do not store the Meter in a place of humidity, 

     high temperature, explosive, inflammable and 

     strong magnetic field.

B. Replacing the Battery
        Warning

To avoid false readings, which could lead to possible 

electric shock or personal injury, replace the battery 

as soon as the battery indicator   appears.

To replace the battery:

1. Turn the Meter off. Disconnect the connection 

    between the testing leads and the circuit under test, 

    and remove the testing leads away from the input 

    terminals of the Meter.

2. Remove the screw from battery compartment and 

    separate the battery compartment from the case 

    bottom.

3. Remove the battery from the battery compartment.

4. Replace the battery with a new 3V battery 

5. Rejoin the case bottom and the battery 

    compartment, and reinstall the screw.

UT118A/B

 

Operating Manual

           Development District,Hu Men Town,Dong Guan Cit

y,

           Guang Dong Province,China 

3. AC Voltage Measurement

        Warning

To avoid harm to the Meter, never input higher than 

300V  voltage  although  it  is  possible  to  measure 

higher voltage.

●  Set the switch to 

V     .  

  

●  Press 

SELECT

 to select AC voltage measurement

     mode

●  Connect the test leads across with the object being 

    measured. The measured value shows on the 

    display.

●  When voltage measurement has been completed, 

    disconnect the connection between the testing 

    leads and the circuit under test, and remove the 

    testing leads away from the input terminal of the 

    meter.

Note:

The threshold voltage of AC voltage is around 400mV.

4. EF Measurement( UT118B only)

        Warning

To avoid harm to the Meter, never input higher than 

300V  voltage  although  it  is  possible  to  measure 

higher voltage.

●  Set the switch to 

V     EF

 and remove the test lead 

    from the input terminal.

●  Press 

SELECT

 to select EF measurement mode.

●  Unconnected measurement. Place the red 

    terminals towards the object being measured. 

5.                     auto measurement

        Warning

To avoid damages to the Meter or to the devices 

under test, disconnect circuit power and discharge all 

the high-voltage capacitors before measurement.

●  Set the switch to                   .   

●  Auto measuremeng mode is a default, under this 

    mode can measure Resistance, Diode, Continuity 

    and Capacitance automatically.

●  For better accuracy, it better separate the object 

    being measured from the circuit before 

    measurement.

●  When voltage measurement has been completed, 

    disconnect the connection between the testing 

    leads and the circuit under test. 

Note:

Under auto measurement mode, when input: 

        Resistance: <15Ω or >10MΩ

        Capacitance: <400pF or >1mF

Will get an irresponsible value.

6. Resistance Measurement

        Warning

To avoid damages to the Meter or to the devices 

under test, disconnect circuit power and discharge 

all  the  high-voltage  capacitors  before  measuring 

resistance.

To measure resistance, do the following:

●  Set he rotary switch to                   .

●  Press 

SELECT

 to select 

Ω

 measurement mode.

●  Connect the test leads across with the object being 

    measured. 

●  When resistance measurement has been 

    completed, disconnect the connection between the 

    testing leads and the circuit under test, and remove 

    the testing leads away from the input terminals of 

    the Meter.

7

. Continuity Test

         Warning

To avoid damages to the Meter or to the devices 

under test, disconnect circuit power and discharge all 

the high-voltage capacitors before continuity test.

To measure resistance, do the following:

●  Set he rotary switch to                    .

●  Press SELECT to select       measurement mode.

●  Connect the test leads across with the object being 

    measured. 

●  The buzzer sounds continuously if the resistance 

    of a circuit under test is≤30Ω, it indicates the circuit 

    is in good connection.

Note:

When continuity measurement has been completed, 

disconnect the connection between the testing leads 

and the circuit under test, and remove the testing 

leads away from the input terminals of the Meter

8. Diodes Test

        Warning

To avoid damages to the Meter or to the devices 

under test, disconnect circuit power and discharge all 

the high-voltage capacitors before measuring diodes.

To measure diode, do the following:

●  Set he rotary switch to                   .

●  Press SELECT to select       measurement mode

●  For better accuracy, it better separate the object 

    being measured from the circuit before 

    measurement.

●  When diodes measurement has been completed, 

    disconnect the connection between the testing 

    leads and the circuit under test. 

9. Capacitance Measurement  

         Warning

To avoid damage to the Meter or to the equipment 

under test, disconnect circuit power and discharge 

all  high-voltage  capacitors  before  measuring 

capacitance.  Use the DC Voltage function to confirm 

that the capacitor is discharged.

To  measure  capacitance,  connect  the  Meter  as 

follows:

●  Set he rotary switch to                   .

●  Press SELECT to select      measurement mode

●  For better accuracy, it better separate the object 

    being measured from the circuit before 

    measurement.

●  When diodes measurement has been completed, 

    disconnect the connection between the testing 

    leads and the circuit under test. 

General Specifications

●  Maximum voltage between red Terminals and 

    Grounding: 300Vrms.

●  Maximum Display: 3000. Updates 4 times/second

●  Temperature:

         Operating: 0

~40

 (32

~104

);   

         Storage: -10

~50

( 14

~122

).

●   Relative Humidity: ≤75% @ 0

~30

                                  ≤ 50% @ 31

~40

●  Altitude:  Operating: 2000m;  Storage: 10000m.

●  Battery Type: 3V Li-MnO2 Button cell battery

●  Battery Deficiency:  Display “         ”.

●  Dimensions (HxWxL):  20.18x26.5x181.5mm

●  Weight:  Approx.90g (battery included).

Accuracy Specifications

Accuracy:±(a% b digits)guarantee for 1 year.

Operating temperature: 18

 ~ 28

.

Relative humidity: <75%.

A . AC Voltage

Range

 

Resolution   Accuracy     

Overload Protection

 3V 

0.001V 

 30V 

0.01V 

 ± (1%+4)          300Vrms 

 300V 

0.1V

 

Remark: Input Impedance: ≥10MΩ

               Frequency Response: 40Hz ~ 400Hz 

 

B. DC Voltage

Ran

ge     Resolution  Accuracy    

Overload Protection

 3V 

0.001V 

 

 30V 

0.01V          ± (1%+3)          300Vrms 

 300V 

0.1V

Remark: Input Impedance: ≥10MΩ  

 

C.  EF Test (UT18B Only)

 Range 

                   Remarks

                             <10mm: buzzer beeps;

 220V/50H           10~50mm: may and may not beeps;

                             >50mm: Buzzer not sound

D. Resistance

Range 

Resolution   Accuracy     

Overload Protection

 300Ω 

0.1Ω 

   

       

 3kΩ 

1Ω 

 

 

 30kΩ 

10Ω 

 

 

 300kΩ 

100Ω 

 

       300Vrms 

 3MΩ 

1kΩ 

 

 30MΩ 

10kΩ

Remark: Under auto scan mode the max range is   

3MΩ.

± (1%+3) 

± (1.5%+5)

±(3%+5)

 300Vrms

0.1 Ω