Note
20
Model UT106: OPERATING MANUAL
The test leads can add 0.1 to 0.2 of error to the
resistance measurement. To obtain precision readings
in low-resistance, that is the range of 200 , short-
circuit the input terminals beforehand and record the
reading obtained (called this reading as X). (X) is the
a d d i t i o n a l r e s i s t a n c e f r o m t h e t e s t l e a d .
Then use the equation:
measured resistance value (Y) – (X) = precision
readings of resistance.
When the resistance reading 0.5 in the short-circuit
condition, please check for loose test leads or other
reasons.
For high resistance (>1M ), it is normal taking several
seconds to obtain a stable reading, and it is better to
choose shorter test lead.
When there is no input, for example in open circuit
condition, the Meter displays
“1”
.
When resistance measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test.