User’s Manual UDP6900 Series Digital Control Power Supply
Instruments.uni-trend.com
24
/
44
The "10 code" mode indicates that start from the start point is the opened state, until the total points
are all set to follow the cycle state of "10";
PS: This template only changes the output state of the point.
2.
Fixed Duration
Fixed duration is used to open/close the fixed open time (maximum 9999.9s) and the close time
(maximum 9999.9s).
“Open time” Ton represents that when the output is turned on, the delayer extends the set time and
then turns to the closed state.
“Close time” Toff represents that when the output is turned off, the delayer extends the set time and
then turns to the opened state.
After custom time mode generates the group data, it defaults to the off output state as the starting
state.
PS: This template only changes time of the point.
3.
Duration Increment
Duration increment is a mode that extends the control output channel to the open or close state by
increasing the step value (Tstep<9999.9s) based on Tbase.
Calculation formula: T(i+1)=T(i)+Tstep,T(0) = Tbase
PS: This template only changes time of the point.
4.
Duration Decrement
Duration decrement is a mode that extends the control output channel to the open or close state by
decreasing the step value (Tstep<9999.9s) based on Tbase.
Calculation formula: T(i+1)=T(i)-Tstep,T(0) = Tbase
PS: This template only changes time of the point.
Storage and Read
After the parameter of the list output and delayer is set by manual or by template settings, the user can use
the storage function to save the parameters in the system or external media for later use.
Storage
This device can separately save 10 group setting values of the three functions of the list output, delayer and
state file, and can also save to the external storager.
Press the Menu key on the panel, and click on the “User Group” button to enter the storage interface.
The storage interface includes the set value of each channel voltage and current, over voltage and over
current.