Figure 10-1 Basic Parameter of Delayer
<Delayer Setup>
Shortcur Key
Setting
Description
FileNum
/
The currently loading list file, press [Save] shortcut key to
enter <Delayer File> to save and recall the file.
Start
0-199
Select the line to start the test from the list on the right side,
the default is 0.
Group
1-199
Count the output step starts from the initial group number,
the default is 1.
Repeat
1-9999
Times of test cycly, the default is 1.
Finish
Stop, Hold
Stop: stop the output when the test is finished.
Hold: output the last test item when the test is finished.
Enable
Off,On
Whether to enable the delayer test, if the delayer test is
enabled, DELA displays in title bar.
After the delayer test is enabled, press [On/Off] key to start
the delayer test and the indicator will blinking to distinguish
from the normal output.
Enable the delayer test will turn off the list function.
Sheet
Enter sheet setting
Memory
Enter disk (local disk and Udisk)
Press [Sheet] key to enter the parameter setup of delayer output.
Figure 10-2 Parameter Setup of Delayer Output.