2. PANEL OPERATION
2-1. <MEAS DISPLAY>
Press [MEAS] key, the <MEAS DISPLAY> page will be displayed on screen as shown below:
< MEAS DISP >
MEAS
FUNC : Cp-D
RANGE : AUTO
DISP
FREQ : 1 kHz
SPEED : SLOW
LEVEL : 1.00V
COMP : OFF
BIN
/
Cp : -0.0015p F
D : 0.3697
DISP
Vm:OFF Im:OFF
QUICK
CLEAR
Use softkeys to select
2-1-1. MEAS DISP
In <MEAS DISP> page, use the four direction keys (▲▼◄►) to move cursor to choose of FUNC, FREQ,
LEVEL, RANGE, SPEED, COMP. Pressing [QUICK CLEAR] softkey can make quick clear to all parameters.
Test parameter
When parameter “Cp-D” of FUNC is selected, other test parameters are display in the right side of the screen.
Each parameter can be selected by its softkey. There are test parameters of Cs-D, Cp-D, Ls-Q, Lp-Q, Rs-Q,
Rs-D, Rp-Q, Rp-D, R-X, Zs-θ, Zs-r.
< MEAS DISP >
Cs-D
FUNC : Cp-D
RANGE : AUTO
FREQ : 1 kHz
SPEED : SLOW
LEVEL : 1.00V
COMP : OFF
Cp-D
/
Cp : -0.0015p F
D : 0.3697
Ls-Q
Lp-Q
Vm:OFF Im:OFF
MORE
1/4
Use softkeys to select