AP301-FULL-HD_rev.09_UK
16
4.2.1.
BATTERY STATUS
In order to know the charge level of the battery and thus the remaining autonomy, look at
the battery icon on the screen, in the upper part on the left.
4.2.1.1.
PREPARING THE INSTRUMENT FOR USE
The instrument switches on in the same mode in use when it was switched off, the only
exceptions being:
•
For safety reasons, output voltage from the input connector
RF IN (5)
is always
excluded.
•
The status of digital boards QPSK , OFDM and QAM is not re-established.
Every time a key is pressed, a short acoustic note confirms this operation.
This acoustic note can be deactivated from the main menu, selecting the item “BUZZER”.
4.3.
AUDIO/VIDEO SETTINGS
In order to adjust volume, brightness, contrast, colour, buzzer volume and tone, press
several times the key
until the desired adjustment appears. Modify its value turning
the
shaft encoder
. Press
ENTER
to confirm it or just wait for time out.
4.4.
TUNING
Tuning can be carried out by frequency, by channel or by program. To select one of these
modes, press the corresponding key:
PR
for program,
CH
for channel,
FR
for frequency to
alternate between channel mode and frequency mode. The indicator corresponding to the
selection flashes on the status bar. Whichever mode is selected, tuning is carried out
turning the shaft encoder or inserting values through the alphanumeric keypad. Press
ENTER
to confirm setting.
In satellite band, the frequency display, from 1
st
IF to transponder frequency, can be
alternated pressing the
LO
key.
Moreover, using the
TP SAT
function in the side menu (which is activated pressing
ENTER
when the instrument
is in satellite band), satellite transponders’ tuning is carried out
selecting the transponder number.
To select the desired satellite, press keys
2nd
and
TP SAT
and select the item SATELLITE
in the menu shown on the screen. Then press
ENTER.
Turn the shaft-encoder to highlight the desired satellite, then press
ENTER
again to confirm
it.
4.4.1.
LOCAL OSCILLATOR (LO) ENGAGED MENU
Keep the
LO (4)
key pressed
for approx. two seconds to activate the related menu:
•
LO DEFAULT: to display the default frequency of local oscillators of a universal LNB:
LO1=9750 MHz, LO2=10600 MHz
•
LO USER: to set the frequency of four available local oscillators.
To exit menu and go back to normal functioning press the
LO (4)
key.