The sensor may be mounted vertically to horizontally, but the sensor tip 
must not be higher than the sensor end!

C

ommissioning

The transmitter housing contains two green terminals.
The UMS oxygen sensor is connected to the 5-row.
Power is supplied via the current loop on the 4-row terminal.
(The power cable must not bend / break the light pipe!) When closing the 
cable clamps, be sure to fix the lock nut, otherwise the clamp will rotate 

and its bonding will come off. This allows water to penetrate into the 
housing.
Screw the housing cover in the correct position, otherwise the LED 

indicator will not work. Cable clamps below, LED indicator on the left, push
button on the right!)
In case the transmitter is connected to a 20 mA input with power supply, 

only the two outer terminals must be connected:

UMS Transmitter

20 mA Input

RED

„Loop(+)“, respectively LOOP

BLUE

„Loop(-)“, respectively

 LOOP RET

2

Summary of Contents for OxyScan Transmitter waterproof

Page 1: ...sor electrochemical Always store the sensor in its calibration chamber when not in use The membrane thin be careful atthe sensor tip can be cleaned carefully with a moistened cotton ball A constantly...

Page 2: ...When closing the cable clamps be sure to fix the lock nut otherwise the clamp will rotate and its bonding will come off This allows water to penetrate into the housing Screw the housing cover in the...

Page 3: ...l operation the LED on the transmitter is shining with light intensity blue or green depending on the measuring parameter Green relative value sat Blue absolute value mg l The output of the oxygen val...

Page 4: ...DO value too high sensor defective or incorrectly wired short circuit Flashing red DO value too low sensor defective or incorrectly wired Flashing blue NTC error short circuit or not connected The sma...

Page 5: ...e sensor must be in the supplied calibration chamber After this waiting period the calibration is started by pressing the button on the transmitter for at least 3 seconds If the calibration is success...

Page 6: ...on optionally mg l or saturation 4 row clamp Normal 4 20 mA Alarm signal 24 mA Power supply 6 32 VDC min 24 mA Storage and operating temperature of the sensor 0 70 C Measuring ranges Parameter Measuri...

Page 7: ...er use and or malfunction of the device We provide a 12 month legal guarantee on the entire measuring system In the guarantee case please return the sensor or measuring instrument with the test certif...