The sensor may be mounted vertically to horizontally, but the sensor tip
must not be higher than the sensor end!
C
ommissioning
The transmitter housing contains two green terminals.
The UMS oxygen sensor is connected to the 5-row.
Power is supplied via the current loop on the 4-row terminal.
(The power cable must not bend / break the light pipe!) When closing the
cable clamps, be sure to fix the lock nut, otherwise the clamp will rotate
and its bonding will come off. This allows water to penetrate into the
housing.
Screw the housing cover in the correct position, otherwise the LED
indicator will not work. Cable clamps below, LED indicator on the left, push
button on the right!)
In case the transmitter is connected to a 20 mA input with power supply,
only the two outer terminals must be connected:
UMS Transmitter
20 mA Input
RED
„Loop(+)“, respectively LOOP
BLUE
„Loop(-)“, respectively
LOOP RET