UMF Medical
www.umfmedical.com
7
3.2.5.4
Install new actuator - install pins, route cord, and
plug cord into Control Box.
3.2.6
Replace hand control.
See section 3.1.8.
3.2.7
If this does not work, contact UMF Medical Customer
Service at (814)266-8726.
3.3 THE TABLE IS NOT MOVING UP AND DOWN. WHAT
SHOULD I DO?
3.3.1
Check to make sure the powered back section is
working properly.
If not, follow instructions in
section 3.1.
3.3.2
Check visible wires for damage.
Inspect all wires that
are readily visible for signs of cuts, tears, or other
damage that could cause a loss of communication
between components.
3.3.3
Check plugs.
Check all plugs and connections to
ensure that all wires are properly connected.
3.3.4
Check for control box activation.
When the hand
control is pressed, the control box should make a light
clicking noise. If the noise is not heard when
activating the up-down button, replace the hand
control (see section 3.1.8)
3.3.5
Replace Control Box.
See section 3.1.7
3.3.6
If this does not work, contact UMF Medical Customer
Service at (814)266-8726.
3.4 ONE OR MORE OF THE ACTUATORS MAKES A GRINDING OR
OTHERWISE LOUD NOISE WHEN ACTIVATED. WHAT SHOULD I
DO?
3.4.1
Check for foreign obstacles.
Make sure that there are
no foreign obstacles in the path of the actuator that
could cause resistance to motion.
3.4.2
Replace the appropriate actuator.
3.4.2.1
Replace the back actuator.
See section 3.2.5.
3.4.2.2
Replace a base actuator.
3.4.2.2.1
Disconnect power to the table.
3.4.2.2.2
Unplug appropriate actuator cords from the
Control Box and free the cords from the table.
3.4.2.2.3
Remove the back actuator. See section 3.2.5
3.4.2.2.4
Remove the side drawer.
3.4.2.2.5
Remove the four bolts on the underside of the
table and then lift off the upholstered top.
Actuator
Cords
Bolt
locations
Bolt
locations