11
1.1.3 Warning explosion
Cause
Preventive Procedure
Pump explosion is consequence of
excessive pressure inside the pump.
⇒
The maximum pressure inside the pump is
0.03 MPa (GAUGE).
If inlet pressure of the pump is higher than 0.03
MPa (
GAUGE
), eliminate any objects which may
cause obstruct gas flowing to the inlet port.
When using oil mist trap, exchange or clean
them to avoid gas flowing.
1.1.4 High temperature
Cause
Preventive Procedure
Burn at high temperature.
⇒
① Temperature of the pump is high as shown
below
pump body → 70 ~ 80℃
motor → 70 ~ 80℃
②
High temperature surface of the pump may
cause
to burn as a result of direct skin contact
by accident. Do not touch the pump body
during operation. Inspection should be carried
out after stopping and cooling down the
pump.