background image

Vascular flow  
detection 

The 

PD1 combi 

PD1+ combi 

can be  used to detect both surface 

vessels, deeper arteries and veins using either the 5MHz or 8MHz 
transducers. 
To obtain the best signal, apply a liberal amount of coupling gel to 
the area of the vein or artery under investigation. Tilt the transducer 
at approximately 45 degrees to the vessel. Arteries give a high 
pitched pulsatile sound, with veins giving a sound like a roaring 
wind. The optional headset helps to eliminate ambient noise, making 
it easier to detect the weaker signals. 

It is also usual for the 

PD1 combi 

PD1+ combi 

to be used in 

association with a pressure cuff and sphygmomanometer to indicate 
the location and extent of arterial occlusion in the form of ankle/
brachial pressure index and segmental pressures. 
Due to the variation of leg blood pressure over a wide range with the 
systemic pressure, the actual values are less useful than the 
pressure index, which relates the ankle pressure to the pressure 
obtained at the brachial artery. Using the 

PD1 combi 

PD1+ combi 

to measure both pressures will ensure compatibility. In cases where 
patients have peripheral arterial disease using the 

PD1 combi 

PD1+ combi 

, due to its high sensitivity, can be the only technique 

suitable for the measurement of leg blood pressure. 
 
Pressure Index =        Ankle systolic pressure 
                                 Brachial systolic pressure 

 

Normal - ankle systolic pressure > brachial pressure. 
                                 Normal pressure index >1 
                                 Abnormal pressure index <1 
 

Summary of Contents for PD1 combi

Page 1: ...PD1 combi PD1 combi operating instructions ultrasound technologies...

Page 2: ......

Page 3: ...l fetal heart rate detection and rate display on an LCD display and an RS232 data port for the transfer of data to a PC to review the fetal heart rate traces The instrument is supplied complete with t...

Page 4: ...ocated on the front of the PD1 combi unit B The PD1 combi will stay on for approximately 5 minutes or until the on off switch is pressed again With the unit on the volume can be adjusted by the rotary...

Page 5: ...he LCD displays battery condition and fetal heart rate F A battery icon C is displayed when the battery requires changing The fetal pulse icon G flashes at approximately the same rate as the detected...

Page 6: ...s Later on in pregnancy the best signals are generally found higher up the abdomen Avoid sliding the transducer over the abdomen as this results in an increase in the background noise and makes it mor...

Page 7: ...cuff and sphygmomanometer to indicate the location and extent of arterial occlusion in the form of ankle brachial pressure index and segmental pressures Due to the variation of leg blood pressure over...

Page 8: ...ator 50 210 bpm Unit Controls Keys 1 key for unit on off 1 key for unit on off Controls Rotary Volume Rotary Volume Indicators Yellow LED for battery low LCD Display with icon for battery low and puls...

Page 9: ...no audio signal is heard in the loudspeaker consult your supplier When contacting your supplier with a problem please have available the instrument type and serial number The serial number can be fou...

Page 10: ...mise the EM emissions that may be produced by this equipment However in the unlikely event that the unit causes an EM disturbance to adjacent equipment we suggest that the procedure is carried out out...

Page 11: ......

Reviews: