Ultimate Healthcare Heritage II Bariatric User Manual Download Page 14

 

 

Page 

14

 of 

28

 

Issue: 5 - 12/12/2018 
 

8.5  Function mode switch 

 
 

8.5.1  Alternate 

 

 

Alternating

 mode is the default mode for the system. Within this mode the mattress will 

operate in an alternating 1-in-2 cell cycle. The alternating cycle will continue at the 
selected cycle time until another mode is selected. 

 
 

8.5.2  Static 

 

 

Pressing the 

Function Mode Selection

 button until the 

Static 

indicator illuminates puts 

the system into 

Static

 mode.  

 
Within this mode the mattress will maintain a selected constant pressure. After 30 
minutes the system will automatically revert back to 

Alternating

 mode.  

 
 

8.5.3  Constant Low Pressure (CLP) 

 

 

Pressing the 

Function Mode Selection

 button until the 

CLP

 

indicator illuminates puts 

the system into 

Constant Low Pressure (CLP)

 mode.  

 
Within this mode the internal cell pressures within the mattress will be reduced by 50% 
and the mattress will provide a comfortable static surface. 

8.6  Auto-Firm 

 

 

Auto-Firm

 mode can be selected by pressing the 

Auto-Firm

 button. In this mode all cells 

inflate to a single pressure setting to provide a firm and stable surface for nursing 
procedures or for patient ingress/egress. The system will automatically return to 

Alternating

 mode at the previously selected comfort level after 20 minutes. 

 

8.7  Alternate cycle time selection  

 

 

The alternating cycle times can be selected to provide an individualised care program for 
each patient.  
 
Within 

Alternating

 mode the alternating cycle time can be selected by pressing the 

Cycle

 button. Selections can be made from 10-30 minutes at 5 minute intervals. The 

cycle time will be displayed in the LCD screen. 

 

8.8  Alarm mute  

 

 

The 

Alarm Mute

 button temporarily resets the audible 

Low Pressure/Power

 

Failure/Service

 alarms. Should the situation not be resolved and the fault condition 

continues the alarm will resume notifying carer. 

 

Summary of Contents for Heritage II Bariatric

Page 1: ...Heritage II Bariatric Bariatric Dynamic Mattress Replacement System UPRA4878 2 User Manual Issue 5 12 12 2018 UPRA4878 2 UM 5...

Page 2: ...Page 2 of 28 Issue 5 12 12 2018...

Page 3: ...8 2 Comfort level 13 8 3 Auto detection 13 8 4 Control panel lockout 13 8 5 Function mode switch 14 8 5 1 Alternate 14 8 5 2 Static 14 8 5 3 Constant Low Pressure CLP 14 8 6 Auto Firm 14 8 7 Alternat...

Page 4: ...Page 4 of 28 Issue 5 12 12 2018 16 Service and Maintenance 21 16 1 General 21 16 2 Fuse replacement 21 16 3 Air filter replacement 21 17 EMC Information 22...

Page 5: ...anyone under 12 years of age The mattress is not suitable for use on patients with unstable fractures The mattress is not intended for use by patients with a BMI Body Mass Index of less than 30 3 Abou...

Page 6: ...essure redistribution Mattress Replacement Power Unit CPR Tag Cable Management Fixings Air Pipe Connector 3 x Static Head Cells TPU 11 x Standard Cells TPU Removable Cover 4 x Smaller Lower Legs Cells...

Page 7: ...that requires correct procedures or practises in order to prevent personal injury BF symbol indicates this product is according to degree of protection against electric shock for type BF equipment The...

Page 8: ...he power cable e g by kinking shearing or other mechanical damages The power cable for this product must be unplugged from the mains power outlet socket and suitably stowed before moving cleaning or m...

Page 9: ...A Classification Class II Type BF Warranty 2 years Operation environment 50C to 400C 15 RH 93 RH no condensation Storage environment 25 70 93 RH no condensation Environment pressure 70 kPa 101 3 kPa W...

Page 10: ...d foot end as viewed from the foot of the bed There are security straps fitted to the base cover of the mattress which should be fastened loosely to convenient points on the bedframe patient surface C...

Page 11: ...tor to the power unit and then ensure that the air tubes are free from any kind of obstruction and are not kinked Plug the power cable into a suitable electrical socket and switch ON using mains power...

Page 12: ...ction 6 Auto Firm Indicator 2 Comfort Control Buttons 7 Operate Standby Button Indicators 3 Comfort Setting Indicators 8 Cycle Time Selection Button Alternate 4 Control Panel Lockout Button Indicator...

Page 13: ...Once activated the indicator will flash and the system will start detecting and will set the internal pressure for the appropriately The indicator light will flash until the patient s weight and posi...

Page 14: ...mattress will be reduced by 50 and the mattress will provide a comfortable static surface 8 6 Auto Firm Auto Firm mode can be selected by pressing the Auto Firm button In this mode all cells inflate t...

Page 15: ...click into the OPEN position At the same time disconnect the air pipe connector from the power unit to speed up the air release If re inflating the mattress ensure the dial of the CPR attachment is ro...

Page 16: ...le is unplugged without turning the power unit OFF an audible alarm will be heard and the Power Failure indicator will illuminate The audible alarm may be cancelled by pressing the Alarm Mute button N...

Page 17: ...e 4 sec Operate or Standby Power On Inflation Failure Alarm Inflate Failure 9 Low Pressure REPEAT Cycle 4 sec Operate or Standby Auto Firm Failure Alarm Auto Firm Failure 10 Low Pressure REPEAT Cycle...

Page 18: ...steps have been checked Press Alarm Mute for system to be verified again Power Failure Alarm Indicator is flashing and sounding Check the power unit is connected to the mains power supply and that the...

Page 19: ...disinfecting ensure that only a damp cloth is used Caution After cleaning dry the mattress out of direct exposure to sunlight Caution Using inappropriate detergents or disinfectants and not observing...

Page 20: ...le dried or air dried They may be tumble dried on a low heat for up to 90 minutes Drying temperature must not exceed 400C The mattress cover and mattress must be dry prior to refitting Caution Exceedi...

Page 21: ...plug if there are abrasions or excessive wears 2 Check mattress cover for signs of wear or damage Ensure mattress cover and tubes are stubbed together correctly 3 Check the air hoses for any kink or...

Page 22: ...RF emissions CISPR 11 Group 1 The GD311 401 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equ...

Page 23: ...ne s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short 5 UT 95 dip in 5 UT 95 dip in Mains...

Page 24: ...m fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked...

Page 25: ...er W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 0 1 0 1 2 0 1 2 0 2 3 0 1 0 38 0 3 8 0 7 3 1 1 2 1 2 2...

Page 26: ...Page 26 of 28 Issue 5 12 12 2018...

Page 27: ...Page 27 of 28 Issue 5 12 12 2018...

Page 28: ...ue 5 12 12 2018 Ultimate Healthcare Ltd Calmore Industrial Estate Nutwood Way Totton Southampton Hampshire SO40 3WW Tel 0333 321 8996 Fax 023 8066 2388 Email info ultimatehealthcare co uk www ultimate...

Reviews: