General
Install the AirPuff assembly in state of the standard shock assembly:
1. Remove the standard shock assembly from the Treadmill:
a. Switch the Treadmill OFF and detach the power cord at the rear.
b. Remove the upper lane assembly.
c. Detach the right side grey connector pulling it to the right handling it by the grey part.
d. Lift all the shock assembly up, it will slide up and place it in a safe place.
2. Take the AirPuff assembly from the package and clear it from all the packaging material.
3. Slide the AirPuff assembly in the Treadmill in the same position of the one you just
remove, pay attention to align guide carefully.
4. Connect the grey connector paying attention to the polarisation position, identify the position
first and then push the connector in.
5. Mount the choosen lane assembly.
Connect power cord and switch on the treadmill, and after selecting the right animal on the
display you should see the following icon beside the Ugo Basile logo at the right.
This will ensure you the AirPuff assembly has been recognized by the system.
Please ensure to load the latest firmware version of the treadmill software before installing the
Shock and detection controller for treadmill AirPuff, minimum version required is 1.5.6.0