Instructions for use (rev 031121) - En 

PROGRAMIX 50 - 100 

 

 

UGIN DENTAIRE 

 

 

3.2.

 

CONDITIONS OF USE, TRANSPORT AND STORAGE 

 

Authorised conditions of use 

-

 

Ambient temperature: +5°C to +40°C 

-

 

Humidity range: maximum relative humidity 80% for temperatures until 31°C, without condensation and 
decrease linear until 50% to 40°C, without condensation. 

-

 

Ambient pressure: up to an altitude of 2000 m above sea level. 

 

Authorised conditions of storage 

-

 

Ambient temperature: -20°C to +65°C 

-

 

Humidity range: maximum relative humidity 80% 

-

 

Ambient pressure: up to an altitude of 2000 m above sea level. 

 

Conditions to transport  

-

 

Wait the complete cooling before packing/transporting. 

-

 

Use the original packaging only, with the protective elements. 

-

 

Take hold the furnace by its base. 

-

 

Install  the  furnace  in  the  packaging  with  the  protective  elements.  Also  protect  the  accessories  for 
transportation. 

 

3.3.

 

ACCESSORIES 

 

Accessories delivered with the furnace 

(included in the packaging)

 

 

DESCRIPTION 

CODE UGIN 

Refractory plate to Programix 50 (160x210) 

FCH05THE0044 

Refractory plate to Programix 100 (210x280) 

FCH05THE0037 

 

 

Accessory recommended 

(not delivered with the furnace 

 to order separately)

 

 

DESCRIPTION 

CODE UGIN 

Fumes extractor 

FCH01EXT0001 

 
 

4.

 

INSTALLATION AND START-UP 

4.1.

 

UNPACK 

Unpack PROGRAMIX 50 / 100 and check that it is in perfect order (furnace and accessories). 

Any defects can be reported to the carrier. 

Remember to get the carrier to sign the delivery note. 

NOTE: keep the original packaging to transport the furnace (see chapter 3.2). 

 
 

Summary of Contents for PROGRAMIX 100

Page 1: ...INSTRUCTIONS FOR USE Burnout furnace PROGRAMIX 50 PROGRAMIX 100 R v 031121...

Page 2: ...RATING PARAMETERS 11 5 3 DEPARTURE DELAY 11 5 4 1st 2nd AND 3rd STAGE AND DELAY FINAL 11 5 5 PROGRAM NUMBER DISPLAY 12 5 6 TEMPERATURE TIME DISPLAY 12 5 7 INDICATOR OF THE UNIT DISPLAYED ON THE PANEL...

Page 3: ...rev 031121 En PROGRAMIX 50 100 UGIN DENTAIRE 3 A EQUIPMENT DESCRIPTION Rear panel N DESCRIPTION 1 Chimney outlet 2 Thermal fuse 3 Power socket 4 Fumes extractor connection socket 5 Location of triac...

Page 4: ...Instructions for use rev 031121 En PROGRAMIX 50 100 UGIN DENTAIRE 4 B ELECTRICAL DIAGRAM...

Page 5: ...h high temperatures do not touch Wear suitable gloves for handling cylinders after heating The unit has to be connected to an electrical installation in accordance with the applicable standard in the...

Page 6: ...used and must be disposed of in accordance with applicable regulations The insulating material are composed of high temperature insulation wool alkaline earth silicate a substance not classified as d...

Page 7: ...in constant use 1050 C 1922 F temporary peak 1150 C 2102 F Time programming 0 to 100 hours Heat rise rate regulated position 1 to 15 C min full position 16 C min and more Numbers of programs 30 progra...

Page 8: ...the original packaging only with the protective elements Take hold the furnace by its base Install the furnace in the packaging with the protective elements Also protect the accessories for transport...

Page 9: ...actor installation UGIN DENTAIRE recommends installing a fumes extractor on the PROGRAMIX 50 100 in order to achieve optimal gas evacuation It will be installed in place of the chimney outlet It is re...

Page 10: ...le 2 Program number display 0 to 29 3 Temperature time display 4 Indicator of the unit displayed on 3 temperature time 5 Mode key 6 Forward key 7 Reverse key 8 Start Stop touch control 9 Cycle advance...

Page 11: ...s the heat rise per unit time rate of rise There are three positions STOP the furnace stops heating this is only used when the next stage is at a Iower temperature 1 to 15 C min heating is regulated a...

Page 12: ...stage programs Programs 20 to 29 are a double 6 stage programs The program number cannot be changed while the furnace is in operation the program number blinks key enabled or when the cycle has begun...

Page 13: ...rogram parameters can be entered or modified using the FORWARD REVERSE keys During that time the program number blinks Once the program has been entered pressing the key a second time locks the progra...

Page 14: ...programme n 2 After installing your furnace according to the instructions in chapter 4 3 Start up activate the switch ON OFF located on the right side of the furnace The main display indicates the te...

Page 15: ...en the furnace door briefly after each stage to check on the silver wire Let us assume for example that the wire melted at 950 C In that case the calibration should be modified as follows Press and th...

Page 16: ...filters However very high over voltages may interfere with your programs As a precautionary measure we recommend that you keep a record of your programs on the table prepared for that purpose see cha...

Page 17: ...ide Visual inspection of heating chamber and refractory plate before use If necessary clean with a vacuum cleaner to remove any particles dust do not use compressed air If damaged or cracked do not us...

Page 18: ...panel and check resistance at the muffle with a multimeter Programix 50 24 ohms Programix 100 15 ohms Replace the muffle as indicated in chapter 9 4 2 The furnace does not heat and furnace indicator l...

Page 19: ...tance not classified as dangerous according to Regulation 1272 2008 EC CLP The muffle replacement requires contact with the heating chamber and the insulation plates The personnel involved must be equ...

Page 20: ...RAMIX 50 100 UGIN DENTAIRE 20 10 TABLE OF PROGRAMS 10 1 TABLE OF SINGLE PROGRAMS No 0 to No 19 No Start temp RP1 Temp Hold time RP2 Temp Hold time RP3 Temp Hold time Final temp Description 0 1 2 3 4 5...

Page 21: ...e rev 031121 En PROGRAMIX 50 100 UGIN DENTAIRE 21 10 2 TABLE OF DOUBLE PROGRAMS No 20 to No 29 No Start temp RP1 Temp Hold time RP2 Temp Hold time RP3 Temp Hold time Final temp Description 20 21 22 23...

Page 22: ......

Reviews: