SERVICING AND PRECAUTIONS FOR USE 

 

SERVICING 

 

Except for cleaning the machine outside using products without solvents, the DUCATRON Quattro needs 
no particular servicing operation. 
However  it  is  recommended  that  the  cooling  circuit  reservoir  water  level  is  checked  about  every  three 
months. 
  
To carry out this operation:  

 

-

 

Cut off power to the machine (isolating switch on right to "OFF"). 

-

 

Take out front lower panel without forgetting to disconnect earth lead. 

-

 

Undo band that secures cover assembly. 

-

 

Release cover to gain access to reservoir opening. 

-

 

Add any needed water so that the water level is about 5 cm below the upper edge. 

-

 

Put back the various parts, checking that the cover band is correctly positioned and locked up. 

-

 

Re-connect panel earth lead. 

-

 

Close up machine (four quarter-turn  four plugs). 

-

 

Reset mains power. 

-

 

The DUCATRON QUATTRO is operational again.  

 
 
PRECAUTIONS FOR USE 

 

Before any use, we would advise you to check that no object or residue is present in the machine arm. 
 
It is essential that the cradle fits perfectly with the cast cylinder for any operation to be good. It is best to 
carry out this check when the machine is cold. 
The machine is delivered with four cylinder holder cradles (DEGUSSA Standard) and two others that can be 
used with 40mm Ø x 55mm long and 90mm Ø x 70mm long cylinders. 
Other  types  can  be  used  on  condition  that  their  dimensions  and  weights  remain  compatible  with  the 
accessories of the machine. 
 
To prevent any risk of alloy projection during centrifugation, several points should be checked: 
 

-

 

Distance between cup lip and cylinder should not exceed 6mm maximum. 

-

 

The amount of alloy should be adapted to the type of cylinder being used. 

-

 

The cylinder should not show any cracks. 

-

 

The size of the crucible lip should be compatible. 

-

 

The cast cone should be sufficient. 

 
Because of the high temperature of the cylinder whilst it is being handled (on coming out of the oven) it is 
strongly recommended that you use a gripper and suitable gloves. 
 
 
 
 

Summary of Contents for DUCATRON QUATTRO

Page 1: ...TIONS FOR USE ISTRUZIONI D USO MODO DE EMPLEO DUCATRON QUATTRO 25 rue de la Tuilerie 38170 SEYSSINET PARISET France T l National 04 76 84 45 45 T l International 33 4 76 84 45 43 Email info ugin denta...

Page 2: ...Rev 080119 DUCATRON QUATTRO MODE D EMPLOI 1 INSTRUCTIONS FOR USE 13 MANUALE DI ISTRUZIONE 25 MANUAL DE INSTRUCCIONES 37 49...

Page 3: ...ISTIQUES TECHNIQUES 3 INSTALLATION ET MISE EN SERVICE 4 DESCRIPTION DU TABLEAU DE COMMANDE 5 ESSAI EN FONCTIONNEMENT A VIDE 6 EQUILIBRAGE DU BRAS 6 UTILISATION DES CREUSETS 7 DEROULEMENT D UN CYCLE 8...

Page 4: ...nneau de commande signale le manque d eau dans le circuit de refroidissement D s qu il est actif la chauffe devient inop rante Le disfonctionnement ventuel du g n rateur est lui aussi indiqu par un vo...

Page 5: ...ges non pr cieux semi pr cieux et pr cieux cylindres 6X et 9X TEMPS MOYEN DE FUSION POUR 25g D ALLIAGE NEUF 1 minute TEMOIN D ALERTE NIVEAU D EAU circuit de refroidissement TEMOIN D ALERTE NON FONCTIO...

Page 6: ...la machine lors de la centrifugation 3 Retirer les bouchons puis desserrer les 4 vis du panneau avant de la machine l aide d un tournevis cruciforme D poser le panneau et d connecter le fil de terre...

Page 7: ...e 7 Voyant de d marrage rapide 8 Voyants niveau de puissance g n rateur de 20 100 9 Touche de s lection de puissance 10 Voyant FUSION 11 Voyant INJECTION 12 Touche STOP commande l arr t du cycle ou de...

Page 8: ...actionner la touche INJECT 10 Apr s une quinzaine de secondes utiliser la touche STOP pour arr ter le cycle 11 Le capot se d verrouille automatiquement quelques secondes plus tard EQUILIBRAGE DU BRAS...

Page 9: ...elle est employ e avec le creuset silice Enti rement lisse elle prot ge l alliage de toute contamination par les composants du gobelet ATTENTION Ne jamais utiliser la gaine en carbone vitrifi pour la...

Page 10: ...he SELF reste inop rante A l aide de la fourche de pr hension placer l ensemble cylindre berceau dans le compartiment du bras Activer le mouvement de la self de fusion touche SELF Deux bips confirment...

Page 11: ...ger la self Un signal sonore permanent signale le d faut bip bip bip Remplissage du r servoir voir chapitre entretien et pr cautions d emploi Si le d faut persiste contacter le SAV UGIN DENTAIRE DEFAU...

Page 12: ...ion secteur La DUCATRON QUATTRO est de nouveau op rationnelle PRECAUTIONS D EMPLOI Avant chaque utilisation nous vous conseillons de v rifier qu aucun objet ou r sidu ne sont pr sents sur le bras de l...

Page 13: ...st tout aussi efficace au travers du hublot trait sp cialement cet effet En cas de n cessit utiliser la baguette quartz et seulement elle pour pousser les plots non fondus et homog n iser ainsi le m l...

Page 14: ...UGIN DENTAIRE DUCATRON QUATTRO Rev 080119 12...

Page 15: ...RODUCTION 14 TECHNICAL CHARACTERISTICS 15 INSTALLATION AND START UP 16 CONTROL PANEL DESCRIPTION 17 OPERATIONAL TEST WHEN EMPTY 18 ARM BALANCING 18 USING CRUCIBLES 19 CYCLE SEQUENCE 20 OPERATIONAL FAU...

Page 16: ...icates any lack of water in the cooling circuit to protect the self induction system Then the light is on the heating system becomes inoperable A warning light on the front face of the machine indicat...

Page 17: ...and 3X Rapid for non precious semi precious and precious alloys cylinders 6X and 9X MEAN FUSION TIME FOR 25g NEW ALLOY 1 minute WATER LEVEL WARNING LIGHT Cooling circuit GENERATOR MALFUNCTION WARNING...

Page 18: ...tability brings about vibrations and considerably increases the machine sound level during any centrifuge operation 3 Take out plugs and then unscrew the four machine panel screws using a crosshead sc...

Page 19: ...ode selector switch 7 Rapid start warning light 8 Generator power level warning lights from 20 to 100 9 Power selection switch 10 FUSION warning light 11 INJECTION warning light 12 STOP button cycle s...

Page 20: ...ECT switch 10 After some fifteen seconds push on STOP button to stop cycle 11 The cover automatically unlocks some seconds later ARM BALANCING The DUCATRON Quattro arm balancing system is identical to...

Page 21: ...using precious or semi precious alloys it can be used with of silica crucible Being smooth it protects the alloy from any contamination by cup components PLEASE NOTE Never use the vitrified carbon sl...

Page 22: ...en the sound signal and the warning light are not activated the SELF coil button remains inoperative Use the gripper fork to place the cylinder and cradle assembly in the arm compartments Activate the...

Page 23: ...the fault beep beep beep Filling reservoir See section on servicing and precautions on use If fault persists contact the after sales service UGIN DENTAIRE GENERATOR FAULT Generator fault warning ligh...

Page 24: ...RECAUTIONS FOR USE Before any use we would advise you to check that no object or residue is present in the machine arm It is essential that the cradle fits perfectly with the cast cylinder for any ope...

Page 25: ...is perfectly adequate using the heat treated porthole window specially designed for this purpose Use the quartz baton and only the quartz baton to push around the unmelted slugs and to homogenize the...

Reviews: