VACUUM PUMP 

 

 

UGIN DENTAIRE 

 

B.

 

INSTRUCTIONS FOR USE

 

1.

 

SAFETY INSTRUCTIONS 
1.1.

 

USE 

VACUUM PUMP is a hermetic group without oil bath, with non-return valve. It is recommended to be used 
with UGIN DENTAIRE ceramic and pressing furnaces. This equipment has to be operated only by dental 
professionals

VACUUM PUMP has to be used in the conditions and for the intended use in this manual

. Any other use 

that will be made of it is considered inappropriate. In this case, the manufacturer declines all liability and 
immediately cancels the equipment warranty. 
According  to  the  pr

oduct’s  instructions,  it  is  imperative  to 

read  carefully  the  manual

  before  furnace 

installation and operations. Retain the manual for future reference. 

1.2.

 

SAFETY INSTRUCTIONS 

Pictograms’ meaning used in these instructions for use:

 

 

 

General warning

 

 

 

Electrical hazard

 

 

 

Burn hazard - Hot surface

 

 

Observe following safety instructions: 

 

-

 

The unit has to be 

connected to an electrical installation in accordance

 with the applicable 

standard in the country in which it is being used. This installation has to provide 

protection 

against overcurrent, overload and ground faults

-

 

Connect it directly to socket

. Do not use multiple sockets. 

-

 

Ensure that the 

electric cord

 is in perfect condition to prevent short circuits. 

-

 

Before any 

technical intervention

,  the  unit  has  to be  disconnected from  the mains power 

supply. 

-

 

Do not spill liquid

 on the ventilation openings or inside the equipment. 

 

-

 

During operation and after use, 

the vacuum pump surface may reach high temperatures: do 

not touch.

 

 

 

-

 

This vacuum pump was designed to pumping non-aggressive gases containing oxygen less than 
22%. 

Use only with UGIN DENTAIRE ceramic and pressing furnaces. 

-

 

Never use this pump to suck up corrosive liquids or water vapour. 

 

-

 

Use only 

genuine spare parts

. The use of non-original spare parts voids any warranty for your 

device. 

-

 

Do not introduce stranger objects inside

 the unit during the maintenance operations.  It is 

forbidden to modify the material without authorization. 

-

 

Do not execute different operations of maintenance

 from those brought back in the manual. 

Whichever operation not included in this manual, can involve risks. 

-

 

For  whichever  information  concerning  installation,  maintenance  and  use,  contact  the 

customer’s service UGIN 

DENTAIRE.

 

 

Summary of Contents for 201221

Page 1: ...INSTRUCTIONS FOR USE VACUUM PUMP R v 201221...

Page 2: ...Y INSTRUCTIONS 3 1 1 USE 3 1 2 SAFETY INSTRUCTIONS 3 1 3 WASTE DISPOSAL 4 2 EC DECLARATION OF CONFORMITY 4 3 TECHNICAL INFORMATION 4 4 INSTALLATION AND START UP 5 4 1 UNPACK 5 4 2 INSTALLATION 5 4 3 S...

Page 3: ...221 VACUUM PUMP UGIN DENTAIRE 2 A SPARE PARTS N DESCRIPTION Quantity Code UGIN 7 Teflon cup 2 FCE05RAC0044 11 O ring 2 FCE05RAC0042 12 Valve washer lower 2 FCE05MEC0107 13 Valve 4 FCE05MEC0108 17 Rec...

Page 4: ...ch it is being used This installation has to provide protection against overcurrent overload and ground faults Connect it directly to socket Do not use multiple sockets Ensure that the electric cord i...

Page 5: ...erie 38170 SEYSSINET PARISET FRANCE UGIN DENTAIRE declares on its own exclusive responsibility that the product Oil less vacuum pump Serial number see the equipment s EC declaration of conformity Fulf...

Page 6: ...n the pump must be used in a well ventilated room without humidity and with low levels of dust The room temperature should not exceed 40 C If it is not used keep it in a room without humidity and prot...

Page 7: ...ng vibration during operation 1 Connect the Tricolair tube to the ribbed elbow connector 2 Connect the other end of this tube to the kiln at rear NOTE respect the direction of connection 3 Connect the...

Page 8: ...TROUBLE SHOOTING Problem Cause Procedure to followed Motor doesn t start No power sources Check other probable causes Insufficient voltage Check the voltage Malfunction of capacitor Check capacitor i...

Page 9: ......

Reviews: