INSTRUCTIONS FOR USE
VACUUM PUMP
Rév. 201221
Page 1: ...INSTRUCTIONS FOR USE VACUUM PUMP R v 201221...
Page 2: ...Y INSTRUCTIONS 3 1 1 USE 3 1 2 SAFETY INSTRUCTIONS 3 1 3 WASTE DISPOSAL 4 2 EC DECLARATION OF CONFORMITY 4 3 TECHNICAL INFORMATION 4 4 INSTALLATION AND START UP 5 4 1 UNPACK 5 4 2 INSTALLATION 5 4 3 S...
Page 3: ...221 VACUUM PUMP UGIN DENTAIRE 2 A SPARE PARTS N DESCRIPTION Quantity Code UGIN 7 Teflon cup 2 FCE05RAC0044 11 O ring 2 FCE05RAC0042 12 Valve washer lower 2 FCE05MEC0107 13 Valve 4 FCE05MEC0108 17 Rec...
Page 4: ...ch it is being used This installation has to provide protection against overcurrent overload and ground faults Connect it directly to socket Do not use multiple sockets Ensure that the electric cord i...
Page 5: ...erie 38170 SEYSSINET PARISET FRANCE UGIN DENTAIRE declares on its own exclusive responsibility that the product Oil less vacuum pump Serial number see the equipment s EC declaration of conformity Fulf...
Page 6: ...n the pump must be used in a well ventilated room without humidity and with low levels of dust The room temperature should not exceed 40 C If it is not used keep it in a room without humidity and prot...
Page 7: ...ng vibration during operation 1 Connect the Tricolair tube to the ribbed elbow connector 2 Connect the other end of this tube to the kiln at rear NOTE respect the direction of connection 3 Connect the...
Page 8: ...TROUBLE SHOOTING Problem Cause Procedure to followed Motor doesn t start No power sources Check other probable causes Insufficient voltage Check the voltage Malfunction of capacitor Check capacitor i...