Page 51 of 51
temperature for 100.Press
key to enter inlet water temperature 100
°
C adjustment, press
to adjust temperature for 100,press
key to Complete calibration.
6. Power on for many times,0
°
C:inlet and return water temperature is 0.00
±
0.05,Temperature difference is
0.00
±
0.05. 100
°
C:inlet and return water temperature is 100
±
0.05,Temperature difference is 0.00
±
0.05.
7. After the ice water mixture temperature sensors back to room temperature, put them into a constant
temperature oil/water bath (the temperature is 100
°
C.)
11.5 Installation of RTD Module
Before installation of the RTD module (as shown in the figure below ):
After installation (as shown in the figure below ).
UFM-40
Ultrasonic transit-time flowmeter
T
+31 (0) 165 855 655
E
www.u-f-m.nl
U-F-M b.v.
Argon 24
4751
XC Oud Gastel
The Netherlands