UEi VEI DL259 Instruction Manual Download Page 5

Maintenance

Battery and fuse replacement

Battery – 9V NEDA 1604 or 6LR61
Fuse – AF155, 500mA / 660V 

Specifications

1. AC amps Measurement (45Hz-400Hz

)

Crest Factor  < 3 : 1

2. DC Low amps Measurement

3. DC Volts Measurement

4. AC Volts Measurement (45Hz-2kHz)

Crest Factor  < 3 : 1

5. Ohms Measurement

(Remove screw)

Temperature

adjustment

Range
40A
400A

Resolution
0.01A
0.1A

Accuracy
±3.0%+20dgts
±2.0%+5dgts

Overload Protection
400A Continuous

6. Capacitance Measurement

7. Temperature Measurement

8. Frequency Measurement

9. Continuity Measurement

10. % Duty Cycle

Probe: 

The probes are UL listed to 1000V, 10A, CAT III and are 
double insulated.

Physical Specifications

Range
40µA
400µA

Resolution
0.01µA
0.1µA

Accuracy
±1.0%+5dgts
±1.5%+2dgts

Overload Protection
400µA/600V

Range
4V
40V
400V
1000V

Resolution
1mV
10mV
100mV
1V

Accuracy
±0.9%+2dgts

Overload Protection
600V RMS

Range
4V
40V
400V
1000V

Resolution
1mV
10mV
100mV
1V

Accuracy
±1.9%+3dgts

Overload Protection
600V RMS

Range
4 0 0
4 k
4 0 k
4 0 0 k
4 M
4 0 M

Resolution
10 0 m
1
10
10 0
1 k
10 k

Accuracy
±0.9%+3dgts

±1.2%+3dgts
±1.5%+5dgts

Overload Protection
600V

Range
1µF
10µF
100µF
1000µF
10000µF

Resolution
0.001µF
0.01µF
0.1µF
1µF
1µF

Accuracy
±1.7%+5dgts

±2.5%+15dgts
±15%+100dgts

Overload Protection
600V

Range
999.9Hz
9.999Hz
99.99Hz

Resolution
0.1Hz
1Hz
10Hz

Accuracy
±0.05%+2d

Overload Protection
600V

Open circuit voltage <2.7V
Threshold  Approx:  <50

Overload Protection
600V

DL259-MAN    

P. 4

Range

Resolution

Accuracy

Overload Protection

-40˚ to 68˚F

0.1˚

±5.4˚F

600V

(-40˚ to 20˚C)

(0.1˚)

±(3.0˚C)

68˚ to 400˚F

0.1˚

±1.0% +3.6˚F

(20˚ to 400˚C)

(0.1˚)

(±1.0% +2.0˚C)

400˚ to 2,498˚F

±3% of reading

(400˚ to 1.370˚C)

(1˚)

Range

Resolution

Accuracy

Overload Protection

0.0 - 99.9%

0 .1 %

±0.2% per kHz +0.1 % 6 0 0 V

Operating temperature

32˚ to 104˚F (0˚ to 40˚C)

Storage temperature

-4˚ to 140˚F (-20˚ to 60˚C)

Relative humidity

Maximum relative humidity 80% for 
temperatures up to 31˚C decreasing 
linearly to 50% relative humidity at 40˚C

Altitude

Operating - up to 2000 m
Storage - 10000 m

Pollution degree

2

Installation category

CAT III

Certifications

UL & cUL standard UL 3111-1 listed

Summary of Contents for VEI DL259

Page 1: ...1 800 547 5740 Fax 503 643 6322 www ueitest com email info ueitest com DL259 INSTRUCTION MANUAL ...

Page 2: ...power If an erroneous reading is observed disconnect power immediately and recheck all settings and connections Controls and Indicators 1 Clamp Measure inductive AC current Opens to 1 25 32 mm NOTE The clamp uses a high tension spring to close the jaw Do not allow fingers or objects to become pinched in the base as jaw closes 2 Conductor Alignment Marks Used to aid in the visual alignment of a con...

Page 3: ...in SERIES with the circuit or circuit element under test NEVER CONNECT THE TEST LEADS ACROSS A VOLTAGE SOURCE while the rotary switch is set to the microamps position This can cause damage to the circuit under test or this meter 10 Off Position Turns the meter off Always store your meter in the off position If the meter will not be used for a month or more remove the batteries 11 Display Communica...

Page 4: ...eter will record minimum maximum and elapsed time for all functions except capacitance To activate and review recorded events see figure 1 Figure 1 Capacitance NOTE Some large capacitors take approximately 10 seconds to settle on a value WARNING Capacitors should be completely discharged prior to testing Measure AC or DC Volts Measuring Resistance and continuity Measuring Frequency Duty Cycle Meas...

Page 5: ...d Protection 600V RMS Range 4 0 0 4 k 4 0 k 400k 4 M 4 0 M Resolution 10 0 m 1 10 10 0 1 k 10 k Accuracy 0 9 3dgts 1 2 3dgts 1 5 5dgts Overload Protection 600V Range 1µF 10µF 100µF 1000µF 10000µF Resolution 0 001µF 0 01µF 0 1µF 1µF 1µF Accuracy 1 7 5dgts 2 5 15dgts 15 100dgts Overload Protection 600V Range 999 9Hz 9 999Hz 99 99Hz Resolution 0 1Hz 1Hz 10Hz Accuracy 0 05 2d Overload Protection 600V ...

Page 6: ...to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of original purchase date will be required before warr...

Reviews: