3
LCd dISPLAY
1. Primary LCD: Display T1 (1st thermocouple reading) or T1 - T2
2. Secondary LCD: Display T2
4
OPERATION
•
POWER ON/OFF
Note:
Make sure you have plugged in the thermocouple probe before
turning on.
Press and hold PWR button to turn on the meter. While the meter is
on, press PWR button to turn off the meter.
•
SLEEP MODE (AUTO POWER OFF)
The meter will turn off automatically after 15 minutes if no buttons
are pressed. To disable auto power off function, when the meter
is off, press and hold “
HOLD
” button then press “
PWR
” button
to turn on the meter, the primary LCD shows the following. Sleep
mode is now disabled.
•
TAKING MEASUREMENT
Connect thermocouple. If probe is open or not connected, the LCD
will show ”---”.
•
DIFFERENTIAL (T1 - T2)
Press "
MODE
" to display differential. Press "
MODE
" again to
return to normal display.
• MAX / MIN
Press
MAX / MIN
button to change operation mode from live
data -> MAX -> MIN -> back to live data. Corresponding value will
display on the LCD. These values will reset at power on or when
performing a reset on MAX/MIN.
Reset MAX/MIN
: Press Max/Min button for 2 seconds to reset
MAX/MIN analysis value. The MAX/MIN value is updated to
current reading.
• DATA HOLD
Freeze display of basic measurements or disable. Press HOLD
button again to go back to the temperature reading.
•
CHANGE UNIT
Press
UNIT
button to change unit ˚C or ˚F. The meter default
temperature unit is ˚F. The meter will default to last selected scale.
1
INTROdUCTION
Thank you for purchasing the UEi DTK2 dual input K type
differential thermometer. This thermometer features a built-in
microprocessor and high resolution analog to digital converter
chip. The thermocouple probe responds very quickly to a wide
range of temperature measurements. The industry standard mini
thermocouple jack accepts a wide selection of probe types.
FEATURES
• Dual input with differential (T1 - T2)
• Min / Max capture (T1 & T2)
• Auto power off with disable feature (15 minutes)
• Large LCD screen
• Complete with bead wire temperature probes and 9V battery
• ˚C / ˚F select function
• Magnetic mount
INCLUDES
• Meter
• 9V battery
• Operation manual
• ATT29A K-type bead probe (2)
• Pouch
2
KEYPAd (CONTROLS)
T1
-
T2
HOLD MAX MIN
K
K
2
1
MODE
SET
PWR
MAX
MIN
UNIT
HOLD
PRESS “
UNIT
” TO SELECT ˚C/˚F.
Freeze display
Press “
MODE
” to change operation mode T1 or T1-T2.
The corresponding reading will display on the primary
LCD. The secondary LCD will always display T2.
Press and hold “
PWR
” to turn on the meter or
press it to turn off the meter. The meter will turn off
automatically after 15 minutes if you don’t press any
key. Press “
HOLD
” and “
PWR
” buttons to disable
Auto sleep mode.
Press “
MAX/MIN
” to change operation mode
from live data -> Max.-> Min.-> back to live date. The
corresponding value will display on the LCD. The value
is calculated since powered on or reset.
Press and hold to enter ice-point calibration. (Please
refer to section
5
to ensure correct calibration.)
Press
UNIT
K
K
K
K
Single Input: (Top View)
K
K
T2
K
K
MIN
K
K
MAX
K
K
HOLD
K
K
5
AUTOMATIC ICE-POINT CALIBRATION
The DTK2 has an automatic ice-point calibration function. Please
follow these steps to calibrate your meter to the thermocouple
probes.
• ITEMS NEEDED:
1. DTK2 Thermometer
2. Two k-type probes
3. Crushed ice/water mixture
(distilled water recommended if available)
• CALIBRATION PROCEDURE
1. Connect both probes to the DTK2
2. Place probe tips in the crushed ice/water mixture
3. Power on the DTK2
4. Press and hold SET for five seconds.
a. The meter will enter ice-point calibration and 0.0°C will
flash on both T1 and T2 readings
5. Stir the ice bath with the probes submerged for one minute
6. Continue stirring the ice/water mixture until the readings
return to normal measurement (flashing will stop)
The ice-point calibration is now complete.
DTK2
75.9
°
f
75.9
°
f
T1
K
T1
K
76.1
°
f
76.1
°
f
T2
K
T2
K
Differential Thermometer
WARNING!
Always connect thermocouple before turning the
unit on for most accurate readings.
WARNING!
Make sure to plug sensor with correct polarity.
Do not be alarmed if erratic readings appear on display prior
to normal mode.
MODE
K
K
K
K