background image

MIN/MAX Function

1.  Press           button until “

MIN

” appears in display (Fig 6). Display 

is now showing minimum humidity and temperature reading.

2.  Press           button again until “

MAX

”appears in display (Fig 7). 

Display is now showing maximum recorded humidity and 
temperature readings.

3.  Press           button until “

MIN

” or “

MAX

” disappear from display 

to return to current readings.

4.  Press           button for more than 2 seconds to clear current 

reading from memory.

*

NOTE:

While checking MIN / MAX value for T2, 

T, 

DP, don’t replace the

p robe or the error code will display. The error code will also appear if the
p robe is not installed when checking MIN / MAX values.

T/DP and 

DP/WB

T1               T2

T -  is calculated by subtracting the T2 (Ex. temp.) from T1 (Air temp.).

DP - is calculated by subtracting the DP (dew point) from T2 (Ex. temp.).

1.  Press           button until 

T appears in display (Fig 8).

2.  Press             button until 

DP appears in display (Fig 9).

Low Battery

A battery symbol will display in the upper right corner when the battery
is low. Inaccurate readings will result from low batteries (Fig 10).

1. Open the battery cover on the back of instrument.

2.  Remove the expired batteries.

3.  Insert two new AAA V batteries.

Troubleshooting

Note:

Dry bulb temperature is air (internal) temperature. Dew point

and wet bulb are calculated from internal temperature and humidity.

S p e c i f i c a t i o n s

Standard & Optional Accessories

Optional

External temperature probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ATT31

DTH31-MAN    

P. 3

(Fig 6)

(Fig 7)

(Fig 8)

(Fig 9)

If I See This 

I Should Check

M a l f u n c t i o n

Make sure the time of pressing “        “ - ON/OFF key is more 
than 0.1 sec.
Check that the batteries are inserted correctly.
Replace batteries if needed
Confirm battery indicator is displayed before display 
disappears. If yes, replace with new battery
Check if sleep mode is active. If yes, press “      + 

HOLD

button to disable auto power off function 
Circuit error in RH measurement - Return instrument to UEi
for repair (see warranty)
Circuit error measuring internal temperature - Return 
instrument to UEi for repair (see warranty)
Circuit error in self calibration channel - Return instrument to
UEi for repair (see warranty)
Internal temperature is out of range
External temperature is out of range

No display

Er 1

Er 2

Er 3

Er 4
Er 5

(Fig 10)

Te m p e rature ra n g e

-4 to 122˚F (-20 to +50 ˚ C )

RH %

0 to 100% RH

Wet bulb

-69 to 122˚F (-21.6 to 49.9˚C)

D ew point

-90.4 to 122˚F (-68 to 49.9˚C)

External tempera t u r e

-4 to 158˚F (-20 to 70 ˚ C )

A cc u ra c y

RH %

±3% at 77˚F (25 ˚ C )

Te m p e ra t u r e

±1˚F 0.6˚F (0.6˚C)

Response time

60 seconds ty p i ca l

S i z e

1” (H) x 1 7/8” (W) x 7” (L)

Cable lengt h

45 5/8”

Probe (w/handle)

7 7/8”

B a t t e ry

2 x 1.5V AAA batteries

Summary of Contents for DTH31

Page 1: ...1 800 547 5740 Fax 503 643 6322 www ueitest com email info ueitest com DTH31 INSTRUCTION MANUAL ...

Page 2: ...ulb and dew point measurements Nothing compares to this fully featured hand size digi tal psychrometer Features include Triple display RH dry bulb and wet bulb temp C F selectable Turnable cap to protect sensor T2 Air temp and T2 Dew Point Data hold MIN MAX Automatic power off Safety Notes Before using this instrument read all safety information carefully In this manual the word WARNING is used to...

Page 3: ... in the display Fig 5 Unit will now display external probe temperature NOTE If the external probe isn t plugged into the meter T2 will not appear on display or any delta temperature Operating Instructions WARNING Be sure to open the sensor protection cap before taking measurements Auto Power Off Sleep Function Unit will turn itself off after ten 10 minutes of non use factory programmed The auto po...

Page 4: ... bulb temperature is air internal temperature Dew point and wet bulb are calculated from internal temperature and humidity Specifications Standard Optional Accessories Optional External temperature probe ATT31 DTH31 MAN P 3 Fig 6 Fig 7 Fig 8 Fig 9 If I See This I Should Check Malfunction Make sure the time of pressing ON OFF key is more than 0 1 sec Check that the batteries are inserted correctly ...

Page 5: ...ed to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of original purchase date will be required before w...

Reviews: