AC Amps <400A Jaw

 • AC Amps can be measured in any position of the rotary selector dial.
 • Center wire in guides for best accuracy.
 • Opposing currents cancel each other (use line-splitter when necessary).
 • Keep hands below guard when measuring high current levels.
 • Do not attempt to measure more than 400A AC.

Features:  

    

AC Amps Measurement - Jaw input

Range

Resolution

Accuracy

Overload Protection

60.00A

0.01A

±2.9% + 15dgts

600V RMS

600.0A

0.1A

±1.9% + 8dgts

 

45Hz to 400Hz True RMS

Minimum Current for Clamp Measurement: 0.3A

•  Rotary selector dial = any position
•  Reading show on upper display

• Buzzer sounds at less than < 40Ω.

 

  WARNING  

• Do not measure resistance on a live circuit.

Features:  

  

BT

SELECT

   

Open circuit voltage   < 0.44V

Overload Protection

Threshold Approx. <40Ω

600V RMS

Continuity

•  Default   =  
•  Press SELECT x1 = 

 CAT III

600V                     400A

DL569

V

CATIII    600V

2000µA

CATIV    300V

V

mV

AUTO

•  Default   =  

 CAT III

600V                     400A

DL569

V

CATIII    600V

2000µA

CATIV    300V

V

mV

AUTO

Resistance: < 40kΩ

Features:  

  

BT

SELECT

 

  WARNING  

• Do not measure resistance on a live circuit.

Range

Resolution

Accuracy

Overload Protection

4.000kΩ

0.001kΩ

±1.0% + 4dgts

600V RMS

40.00kΩ

0.01kΩ

k

 CAT III

600V                     400A

DL569

V

CATIII    600V

2000µA

CATIV    300V

V

mV

AUTO

Diode

GOOD DIODE

Reverse Bias

Displays "OL" 

•  Forward voltage drop if forward biased.
•  “O.L.” if reverse biased.

Features:  

    

Diode Test

Range Open Circuit 

Voltage

Test Current 

(Typical)

Overload Protection

4.0V

< 3.0V DC

0.25mA

600V RMS

BAD DIODE

'0' Both directions

or

Open Diode

Displays "OL" 

Both directions

 V

Forward Bias

Displays approx. 

voltage drop

  WARNING  

•   Use CATIII rated test leads or higher.
•   Do not attempt to measure more than 750V AC.
•   Keep hands below line when measuring high current levels.

  WARNING  

 • High Voltage indicator will display and audible alert will sound over 600V AC

 • High Voltage indicator will display (without audible alert) over 30V AC

Features:  

      

Range

Resolution

Accuracy

Overload Protection

400.0V

0.1V

±2.0% + 5dgts

750V RMS

750V

1V

45Hz to 400Hz Averaging RMS 

Voltage: 750V AC

 CAT III

600V                     400A

DL569

V

CATIII    600V

2000µA

CATIV    300V

V

mV

AUTO

High
Voltage
Alert

  WARNING  

•   Use CATIII rated test leads or higher.
•   Do not attempt to measure more than 1000V DC.
•   Keep hands below line when measuring high current levels.

  WARNING  

 • High Voltage indicator will display and audible alert will sound over 600V DC

 • High Voltage indicator will display (without audible alert) over 30V DC

Features:  

  

BT

SELECT

    

Range

Resolution

Accuracy

Overload Protection

400.0mV

0.1mV

±0.5% + 5dgts

1000V RMS

400.0V

0.1V

±0.5% + 5dgts

1000V

1V

±0.8% + 10dgts

Voltage: 1000V DC

 CAT III

600V                     400A

DL569

V

CATIII    600V

2000µA

CATIV    300V

V

mV

AUTO

High
Voltage
Alert

•  Default =       