Non-Contact Voltage
NCV Sensor in the tip.
• Press and hold the NCV button and move the tip of the clamp meter near
voltage source.
• Non-Contact Voltage Detection is used to detect power with sensor located
in the tip of the clamp head, indicates positive response with both an
Audible and Visual alert.
• Do not use Non-contact voltage detector to determine if there is current on
the wire. Detection operation could be affected by socket design, insulation
thickness, type or other factors.
• Voltage indictor light may also light when voltage (>AC/DC 30V) is present
on the meter’s input jack or from an external interference such as motors,
flashlights, etc.
On Voltage
Approx. 24V AC
DL489 True RMS
600V 600A
CAT III
CATII 1000V
CATIII
600V
MAX
30V
T
2
T
1
TEMP
NCV
MAX/
MIN
PEAK
DC
Zero
INRUSH
Temp
/
°F°C
μA
A
V
Hz/D
uty
MFD
RANGE
DC Amps: <600A
• Do not attempt to measure more than 600A DC.
Features:
RANGE
HOLD
MAX/
MIN
PEAK
Range
Resolution
Accuracy
Overload Protection
60A
0.01A
±(2.0% +8 dgts)
600V RMS
600A
0.1A
Minimum current for Clamp measurement: 0.3A
• Default =
(upper)
• Press x1 = (lower)
• Press x2 = (lower)
• Press and hold =
(upper)
µA
µA