background image

S p e c i f i c a t i o n s

Operational Hints

Never

 

apply

 

voltage

 

to

 

the

 

inputs.

Temperatur

affects

 

readings

Allow

 

the

 

CLM100

 

to

 

attain

 

the

ambient

 

temperature

 

of

 

the

 

wire

 

under

 

test.

Use

 

a

 

9

 

Vo

lt 

alkalin

battery

 

only

Use

 

the

 

test

 

leads

 

that

 

came

 

with

 

the

 

unit

 

only.

 

Other

 

leads

will

 

not

 

work.

Ensure

 

that

 

the

 

test

 

leads

 

are

 

clean

 

and

 

in

 

good

 

working

 

order.

Use

 

an

 

abrasive

 

pad

 

similar

 

to

 

the

 

one

 

provided

 

to

 

ensure

 

the

wire

 

under

 

test

 

is

 

clean

 

and

 

free

 

of

 

oxidation.

Ensure

 

that

 

insulation

 

is

 

stripped

 

adequately

 

from

 

both

 

ends

 

of

the

 

wire

 

under

 

test.

 

This

 

will

 

enable

 

measurements

 

to

 

be

 

accurate.

Ensure

 

that

 

the

 

alligator

 

clips

 

are

 

connected

 

across

 

the

 

diameter

of

 

the

 

wire

 

under

 

test.

When

 

measuring

 

wire

 

for

 

insertion

 

into

 

conduit,

 

add

 

extra

 

wire

to

 

compensate

 

for

 

the

 

accuracy

 

of

 

the

 

meter.

 

For

 

example,

 

205

feet

 

of

 

wire

 

is

 

needed

 

to

 

run

 

inside

 

a

 

piece

 

of

 

conduit

 

that

 

is

 

200

feet

 

long.

 

(This

 

would

 

allow

 

2.5

 

feet

 

on

 

each

 

end

 

to

 

attach

 

the

wire.)

 

At

 

205

 

feet

 

and

 

70˚F,

 

the

 

cable

 

length

 

meter

 

has

 

an

accuracy

 

of

 

±7

 

feet.

 

In

 

this

 

case

 

it

 

would

 

be

 

safer

 

to

 

measure

 

out

212

 

feet

 

of

 

wire.

 

This

 

would

 

ensure

 

that

 

enough

 

wire

 

is

 

available

for the 

application.

3.

Strip

 

the

 

insulation

 

back

 

on

 

each

 

end

 

of

 

the

 

wire

 

being

 

tested.

NOTE:

Make sure both ends of the wire under test are clean and

the conductor is fully exposed. The insulation must be stripped
away so the Kelvin clips can be connected across the diameter of
the wire. Use the abrasive pad to clean the wire ends.

4.

Using

 

the

 

selector

 

on

 

the

 

CLM100,

 

turn

 

to

 

the

 

size

 

of

 

wire

under 

test.

5.

Press

 

the

 

“COPPER”

button

 

if

 

copper

 

wire

 

is

 

being

 

tested.

The 

“CU”

enunciator

 

will

 

illuminate

 

in

 

the

 

top

 

left

 

corner

 

of

the 

display.

6.

Press

 

the

 

“ALUM”

button

 

if

 

aluminum

 

wire

 

is

 

being

 

tested.

The

 

“AL”

enunciator

 

will

 

illuminate

 

in

 

the

 

top

 

left

 

corner

 

of

the 

display.

7. Press

 

the

 

“FT”

button

 

if

 

you

 

require

 

readings

 

to

 

be

 

in

 

feet.

The 

“ft”

enunciator

 

will

 

illuminate

 

in

 

the

 

display.

8.

Press

 

the

 

“M”

button

 

if

 

you

 

require

 

readings

 

to

 

be

 

in

 

meters.

The 

“m”

enunciator

 

will

 

illuminate

 

in

 

the

 

display.

9.

Connect

 

a

 

Kelvin

 

clip

 

to

 

one

 

end

 

of

 

the

 

wire

 

and

 

the

 

other

Kelvin

 

clip

 

to

 

the

 

other

 

end

 

of

 

the

 

wire.

10. Read

 

the

 

length

 

of

 

wire

 

directly

 

from

 

the

 

display.

 

Please

 

note

the

 

“k”

enunciator

 

illuminates

 

if

 

measurements

 

are

 

at

 

or

 

above

1

000 

feet/meters.

11. Disconnect

 

the

 

test

 

leads

 

from

 

the

 

unit

 

when

 

not

 

in

 

use.

Low Battery Indication

The

 

“BAT”

 

enunciator

 

will

 

illuminate

 

in

 

the

 

lower

 

left

 

side

 

of

 

the

 

display

to

 

indicate

 

battery

 

voltage

 

is

 

low.

 

The

 

battery

 

should

 

be

 

changed

 

immediately

 

to

 

ensure

 

proper

 

function

 

and

 

accuracy.

 

Only

 

alkaline

 

9

 

Volt

 

batteries

 

should

 

be

 

used.

Measuring Resistance

1. If

 

the

 

resistance

 

to

 

be

 

measured

 

is

 

wire,

 

follow

 

steps

 

1

 

through

3

 

under

 

“Measuring

 

Length

 

of

 

Wire.”

 

Then

 

follow

 

steps

 

2

 

and

 

3

below.

 

If

 

a

 

discrete

 

resistor

 

is

 

being

 

measured,

 

turn

 

the

 

cable

length

 

meter

 

on

 

and

 

perform

 

the

 

c

alibration 

procedure.

2.

Using

 

the

 

selector

 

on

 

the

 

CLM100,

 

turn

 

to

 

“R”

 

on

 

the

 

dial.

3.

Read

 

the

 

resistance

 

of

 

the

 

wire

 

directly

 

from

 

the

 

display.

Measurement 

Range:

Accuracy:

Resistance:

Operating

 

Temperature:

Operating 

Humidity:

Storage

 

Temperature:

Auto

 

power

 

off:

Battery

 

Life:

15

 

to

 

30,000

 

feet

26ga

 

to

 

4/0ga

250MCM

 

to

 

500MCM

±(2%

 

of

 

reading

 

+3

 

LSD)

 

at

 

65˚

 

to

 

75˚F

±(3%

 

of

 

reading

 

+3

 

LSD)

 

below

 

65˚

 

and

 

above

 

75˚F

LSD = least significant digit

0

 

to

 

65

Ω

32˚

 

to

 

104˚F

20%

 

to

 

80%

 

RH

 

to

 

120˚F

After

 

5

 

minutes

 

(approx.)

Depends

 

on

 

use.

 

Typically

 

40

 

hours.

Extended

 

measurements

 

times

 

can

 

lower

battery

 

life

 

to

 

no

 

less

 

than

 

6

 

hours.

CLM100-MAN 

P. 2

Reviews: