19
ANALYZER SPECIFICATIONS
CONT.
Carbon Dioxide resolution is 0.01% below 1% measured value.
Ambient Operating Range
32° to 113°F (0° to 45°C) 10% to 90% RH
Storage Temperature Range
0° to 120°F (-18° to 50°C)
Battery Type / Life
3 AA cells
>8 hours using Alkaline AA cells
Chargers (optional)
100-240v charger, for NiMH batteries only
Dimensions
Weight:
Handset:
Probe:
0 .8kg handset with protective rubber cover
7 .87 x 1 .77 x 3 .5 in (200 x 45 x 90mm)
11 .8 in (300mm) long including handle .
2 .3 dia . x 9 .4 in (6mm dia . x 240mm) long stainless-steel shaft with 6 .5ft
(2m) long neoprene hose . Type K thermocouple
CO Protection Pump:
Operates at 2000ppm measured CO .
The C165 is in conformity with the relevant Union harmonization legislation listed below:
Directive
Title
201430EU
Electromagnectic Capability
201165EU
Restriction of use of certain hazardous substances in electrical
and electronic equipment (RoHS)
The following harmonized standards and technical specifications have been applied:
CERTIFICATION
The UEi C165 is TUV-tested and certified to EN 50379, Parts 1, 2 & 3 in accordance to 1st German Federal Emission Control Ordinance
(BlmSchV)
EMC
EN507270:2015
Safety
EN61010-1:2010
RoHS
IEC62321-2:2013, IEC62321-1:2013; IEC62321-3-1;2013, IEC63321-5:2013, IEC623321-4:2013, IEC62321-7-2:2017, IEC62321-7-1:2015,
IEC62321-6:2015, AHRI 1260
ELECTROMAGNETIC COMPATIBILITY
European Council Directive 89/336/EEC requires electronic equipment not to generate electromagnetic disturbances exceeding defined levels and
have adequate immunity levels for normal operation . Specific standards applicable to this analyzer are stated below .
As there are electrical products in use pre-dating this Directive, they may emit excess electromagnetic radiation levels and, occasionally, it may be
appropriate to check the analyzer before use by:
Use the normal start up sequence in the location where the analyzer will be used .
Switch on all localized electrical equipment capable of causing interference .
Check all readings are as expected . A level of disturbance is acceptable .
If not acceptable, adjust the analyzer’s position to minimize interference or switch off, if possible, the offending equipment during your test .
At the time of writing this manual (November 2020) UEi Test Instruments is not aware of any field-based situation where such interference has
occurred, and this advice is only given to satisfy the requirements of the Directive . This product has been tested for compliance with the following
generic standards and is certified to be compliant .
EN 61000-6-3 : 2011
EN 61000-6-1 : 2007
Specification EC/EMC/KI/C165 details the specific test configuration, performance and conditions of use .