METROLOGICAL INSPECTION
TECHNICAL DETAILS
WARRANTY
DISPOSAL
EXPLANATION OF SYMBOLS
CLEANING
Basically, a metrological inspection is recommended at in-
tervals of 2 years. However, professional users in Germany
have to comply with the aforementioned according to
“Regulation for Operators of Medical Devices”. This can be
implemented either by UEBE Medical GmbH, an authori-
ty responsible for metrology, or authorised maintenance
services. For this, please observe your national provisions.
© 2017 UEBE Medical GmbH. Not to be reproduced, either
in whole or in part. Subject to technical modifications.
7 0830 214 BB 2020-05 EN
Type:
Maximum thermometer, direct
mode
Measuring points/
body reference points:
Rectal, oral, axillary
Range:
32.00 °C to 42.99 °C
Accuracy:
± 0.10 °C between 35.50 °C and
42.00 °C
± 0.20 °C in other temperature
ranges
Operating conditions: Ambient temperature 10 to 40
°C, relative humidity 30 to 85 %
Storage and transport
conditions:
Ambient temperature -10 to 60
°C, relative humidity 25 to 90 %
Battery:
LR41 or SR41, 1.55 V
Battery life:
2 years
IP rating:
IP 22: protection from solid for-
eign bodies with diameters of
12.5 mm and above; protection
against water droplets
Batteries and technical appliances must not be
disposed of with domestic waste, but should be
handed in at the appropriate collection and dis-
posal points.
Clean the thermometer before and after every use, using
a soft cloth and isopropyl alcohol diluted with water, or
cold soapy water.
The device can be immersed in water or disinfectant solu-
tion when cleaning. Do not immerse the device for longer
than 30 minutes. Do not use boiling water, gas or a steam
autoclave to sterilise the device.
This product complies with Council Directive
93/42/EEC of 5 September 2007 concerning
medical devices and bears the mark, CE 0123
(TÜV SÜD Product Service GmbH).
Degree of protection against electric shock:
TYP BF
Please observe operating instructions
Lot number
Manufacturer
During the warranty period of 2 years from the date of
purchase, we remedy defects at our discretion and at our
expense in our plant through repair work or replacement
delivery of a defect-free device. The cost of returning the
device to our factory shall be borne by the sender. Com-
plaints that are sent back without prepayment will not be
accepted by UEBE.
The warranty does not cover normal wear and tear of
wearing parts or damage caused by failure to observe
the instructions for use, improper handling (e.g. breakage,
leaking batteries) or disassembly of the device by the buy-
er. Furthermore, the warranty does not constitute grounds
for asserting claims for damages against us.
Warranty claims can only be advanced in the warranty
period and by presenting proof of purchase. In the event
of a warranty claim, the unit must be sent to the follow-
ing address together with the proof of purchase and a
description of the complaint: UEBE Medical GmbH, Zum
Schlag 18, 97877 Wertheim, Germany.
The statutory claims and rights of the buyer against the
seller (for example, defect claims, producer liability) are
not restricted by this warranty.
UEBE Medical GmbH
Bürgermeister-Kuhn-Straße 22
97900 Külsheim
Germany
www.uebe.com
ERROR MESSAGES
Measured temperature is below 32.00°C and
is therefore outside the measuring range.
Measured temperature is above 42.99°C and
is therefore outside the measuring range.
Electronic fault. If the fault occurs again,
contact the UEBE customer service centre.
0123