![UE Systems Ultraprobe 100 Instruction Manual Download Page 9](http://html.mh-extra.com/html/ue-systems/ultraprobe-100/ultraprobe-100_instruction-manual_834387009.webp)
Version 1
9
To use the WARBLE TONE GENERATOR:
1. Turn Tone Generator on by selecting either "LOW" for a low amplitude signal (usually
recommended for small containers) or "HIGH" for high amplitude. In high, the Warble Tone
Generator will cover up to 113m³ (4,000 cubic feet ) of unobstructed space. When the Tone
Generator is on, a red light (located below the recharge jack in the front) flickers.
2. Place the Warble Tone Generator within the test item/container and seal or close it. Then scan
the suspect areas with the Scanning Module in the Ultraprobe and listen for where the "warble"
ultrasound penetrates. As an example, if the item to be tested is the seal around a window,
place the Warble Tone Generator on one side of the window, close it and proceed to scan on
the opposite side.
To test the condition of the Warble Tone Generator battery, set to the LOW INTENSITY position and
listen to the sound through the Ultraprobe headphones. A smooth continuous warbling sound should
be heard.
If a "beeping" is heard instead, then a full recharge of the Warble Tone Generator is
indicated.
To charge the Warble Tone Generator:
1. Use the recharger.
2. Plug the recharger cord into the recharge jack located on top of the front panel.
3. Plug the recharger into the local current supply
4. A complete charge will take 7 hours.
5. Since there is no memory problem, the Tone Generator may be charged after short intervals
of use.
wtg1 warble tone generator (optional)