SWIFT TEST KIT 

TEK-4

TABLE OF CONTENTS

uclswiftna.com

SECTION 

ONE

:

 LSSM 

|

 LASER LIGHT SOURCE 

|

 QUICK START GUIDE

SECTION 

TWO

:

 LSSM 

|

 LASER LIGHT SOURCE 

|

 TECHNICAL GUIDE

SECTION 

THREE

:

 OPM1 

|

 OPTICAL POWER METER 

|

 QUICK START GUIDE

SECTION 

FOUR

:

 OPM1 

|

 OPTICAL POWER METER 

|

 REFERENCE & TEST GUIDE

TRAINING MANUAL

Summary of Contents for TEK-4

Page 1: ...uclswiftna com SECTION ONE LSSM LASER LIGHT SOURCE QUICK START GUIDE SECTION TWO LSSM LASER LIGHT SOURCE TECHNICAL GUIDE SECTION THREE OPM1 OPTICAL POWER METER QUICK START GUIDE SECTION FOUR OPM1 OPTI...

Page 2: ...SECTION ONE LSSM LASER LIGHT SOURCE QUICK START GUIDE NORTH AMERICA...

Page 3: ...SWIFT This unit offers 1310nm and 1550nm wavelengths When the LSSM is turned on the red light indicates which wavelength is selected The red light blinks to indicate when the 2kHz modulation is turne...

Page 4: ...ts either the 1310nm or 1550nm wavelength on singlemode fiber It is important to use a test jumper to mate the LSSM to the fiber under test TRANSMITTING LIGHT Test Jumper Signal Propagates Down Fiber...

Page 5: ...inserted they can cause damage to the LSSM output port and the unit may need to be repaired OUTPUT PORT MAINTENANCE LSSM When the LSSM s 2kHz modulation is active the selected wavelength indicator wi...

Page 6: ...The LSSM provides over 1000 operations under normal conditions When the LOW BAT LED indicator is illuminated on the device the CR2 battery should be replaced or recharged depending on the battery type...

Page 7: ...long term 15 min warm up 5nm 5nm SC FC ST LC interchangeable 2kHz Push button toggle auto off CR2 10 to 50 C 30 to 60 C 6 1 x 0 94 x 0 75 15 5 x 2 83 x 1 9cm 3oz 85 4g 3330 Earhart Drive Suite 208 Car...

Page 8: ...FCC Rules These limits are designed to provide reasonable protection against harmful inter ference in a residential installation This equipment generates uses and can radiate radio frequency energy a...

Page 9: ...SECTION TWO LSSM LASER LIGHT SOURCE TECHNICAL GUIDE NORTH AMERICA...

Page 10: ...our FI2K Easy removal of the included SC adapter allows the light source ferrule to be cleaned preventing contamination of test jumpers Additional adapters are available This device maintains full co...

Page 11: ...21 CFR 1040 11 Push button toggle auto off CR2 10 to 50 C 30 to 60 C 6 1 x 0 94 x 0 75 15 5 x 2 83 x 1 9cm 3oz 85 4g Interchangable Adapter SC shown Low Battery Indicator Modulation Control 2kHz CW Wa...

Page 12: ...SECTION THREE OPM1 OPTICAL POWER METER QUICK START GUIDE NORTH AMERICA...

Page 13: ...ed to internal storage The display flashes the indexed location where the reading has been saved Toggles through the saved readings on the device The indexed location wavelength and reading will flash...

Page 14: ...ill briefly appear on screen every 8 seconds Press the button to toggle between dB and dBm modes Hold the button down when in dBm mode to set a reference on the OPM1 Press the button once to turn the...

Page 15: ...S Press and hold the Save and Wavelength buttons simultaneously to delete all saved readings in the OPM1 s internal memory The screen flashes 0000 to indicate that the memory is empty NON VOLATILE MEM...

Page 16: ...ckaging is damaged or otherwise unsatisfactory LOW BATTERY The OPM1 provides over 1000 operations under normal conditions When the BATT indicator is shown on the device screen the CR2 battery should b...

Page 17: ...ailable LCD 2kHz 4000 measurements Mini USB live readings or data dump Push button toggle auto off 10 to 55 C 30 to 70 C CR2 6 1 x 0 94 x 0 75 15 5 x 2 83 x 1 9cm 3oz 85 4g 3330 Earhart Drive Suite 20...

Page 18: ...FCC Rules These limits are designed to provide reasonable protection against harmful inter ference in a residential installation This equipment generates uses and can radiate radio frequency energy a...

Page 19: ...SECTION FOUR OPM1 OPTICAL POWER METER REFERENCE TEST GUIDE NORTH AMERICA...

Page 20: ...ch must be taken with OPM1 optical power meter to perform insertion loss testing on a variety of fiber types and setups INTRODUCTION EXAMPLE 1 Singlemode Fiber Point to Point 1 Jumper Page 2 EXAMPLE 2...

Page 21: ...v ada The fiber under test is singlemode with LC connectors on each end The fiber under test may be up to 250 kilometers long In this example the fiber under test is 400 meters long TEST JUMPER The te...

Page 22: ...d of the OPM1 Between 5 00 and 7 00dBm Page 4 The connectors on the Test Jumper should be cleaned to appropriate standards Use an inspection scope to verify fiber cleanliness Plug the connectors on th...

Page 23: ...e with LC connectors on each end This method allows testing of connectors inside bulkheads One or both ends of this fiber under test may be inside bulkheads The fiber under test may be up to 250 kilom...

Page 24: ...tween 5 00 and 7 00dBm Page 4 The connectors on the Test Jumper should be cleaned to appropriate standards Use an inspection scope to verify fiber cleanliness Plug the connectors on the Test Jumper in...

Page 25: ...nectors on each end The fiber under test may be up to 250 kilometers long In this example the fiber under test is 400 meters long TEST JUMPER The test jumper is paired for loopback testing One end has...

Page 26: ...ll display 0 00dB with some minor variation DO NOT switch the RP 460 back to the dBm mode The Test Jumper MUST remain attached to the DLS 355 Do not remove the Test Jumper from the optical link The co...

Page 27: ...h end In this example the fiber under test is 400 meters long TEST JUMPER The test jumper is a one meter multimode fiber with SC and LC connectors The SC connector will be plugged into the LSMM LED so...

Page 28: ...er on the OPM1 The Insertion Loss of the fiber under test will be shown on the OPM1 screen LOSS TEST The connectors on the Test Jumper should be cleaned to appropriate standards Use an inspection scop...

Page 29: ...xample the fiber under test is 400 meters long TEST JUMPER The test jumper is paired for loopback testing One end has an SC connector and an LC connector and the other end has two LC connectors clippe...

Page 30: ...cleaned to appropriate standards Use an inspection scope to verify fiber cleanliness Plug the clipped LC end of the Test Jumper into one end of the Fiber Under Test using the bulkhead as shown Plug th...