background image

Manual LCC RESPIRA 1 

 

 

 

LCC Respira 1 is a Low Cost & Comfortable Powered Air Purifying 

Respirator,  designed  as  respiratory  Personal  Protective  Equipment  (PPE) 
for  medical staff  dealing with  highly infectious disease. As an air-purifying 
respirator  (APR)  it  can  also  be  used  in  other  respiratory-hazardous 
condition. 

 

The  equipment  meets  the  requirements  and  testing  by  BPFK, 

Indonesian Ministry of Health. 

  

A. Components and Functions 

The  equipment  is  made  of  a  mask  mask,  connected  by  inhalation  hose 
(CPAP), to the filter and electronic section. 

LCC-Respira is equipped with HEPA filter (can be replaces with other filter 
in accordance to the use requirement), able to sieve and filter micro particles 
and  virus,  without  stifling  or  suffocating  the  user.  The  filter  is  replacable 
after certain length of usage, ensuring safety in long term usage. Clean air is 
pumped to the mask to ease breathing,  

The  electronic  section  employs  an  electromotor  to  pump  the  filtered  air. 
Electric  controller  is  embedded  to  regulate  the  blower  rate,  according  to 
adjustable settings. On intermediate flow, battery powering the motor can 
last up to 8 hours.  

(a)

                                     (b) 

 

 

Figure 1. (a) Components, (b) Electronics & battery container 

Elektronics & battery 

Blower 

HEPA Filter 

CPAP hose & mask 

Summary of Contents for LCC RESPIRA 1

Page 1: ...Manual LCC RESPIRA 1 Low Cost Comfortable Powered Air Purifying Respirator Hak cipta terdaftar 2020 ...

Page 2: ... menyaring partikel mikro termasuk virus dan tidak pengap bagi pengguna Alat ini aman untuk digunakan oleh tenaga medis karena pada bagian masker dilengkapi lapisan filter yang dapat diganti ganti Udara bersih dipompakan ke dalam masker untuk memudahkan pernapasan Bagian elektronik terdiri dari motor listrik sederhana yang memungkinkan udara dari luar dipompa dan difilter Kontrol elektrik ditanam ...

Page 3: ...Gambar 2 a 2 Pasangkan filter pada Hepa Filter Gambar 2 b kemudian tutup kembali sampai tertutup sempurna Gambar 2 c a b c d Gambar 2 a Cara membuka Hepa filter b Cara Pemasangan Hepa Filter c Blower dan mesin d Selang CPAP dan Masker 3 Sambungkan ujung selang CACP dan masker Gambar 2 d ke blower dan mesin Gambar 2 c hingga seluruhnya terhubung Gambar 3 a ...

Page 4: ...pira adalah 8 jam Lakukan pengisian ulang setelah penggunaan 3 Pastikan seluruh komponen telah tersambung dengan sempurna 4 Nyalakan tombol on off pada bagian elektronik Gambar 4 Gambar 4 Tombol on off dan lampu indikator baterai 5 Setelah mesin menyala udara akan difilter melalui Hepa Filter 6 Udara bersih yang berasal dari Hepa Filter dialirkan ke selang CPAP 7 Selang CPAP mengalirkan udara bers...

Page 5: ...hal ini terjadi lekas ganti baterai Anda dengan cara 1 Buka Kotak hitam mesin 2 Keluarkan baterai pada mesin 3 Ganti baterai dengan baterai yang baru 4 Tutup kembali kotak mesin 5 Isi ulang baterai charging 6 Jika lampu indikator baterai telah terisi penuh maka mesin siap digunakan Gambar 5 Cara penggantian baterai D Langkah Penggunaan LCC Respira 1 Pelindung Muka Masker Filter blower bagian elekt...

Page 6: ...gan mengaitkan head loop 5 Pasang pelindung muka dengan mengaitkan head loop Ukuran headloop dapat disesuaikan dengan ukuran kepala 6 Lalu bernafaslah seperti biasa Tabel 1 Spesifikasi komponen LCC Respira 1 Komponen Spesifikasi Air flow Adjustable 7 15 LPM Filter bakteri HEPA Blower material anti bacterial PLA Berat komponen masker 150 gr Berat kotak filter dan elektronik 480 gr Baterai Rechargea...

Page 7: ...Manual LCC RESPIRA 1 6 Manual LCC RESPIRA 1 English version Low Cost Comfortable Powered Air Purifying Respirator Copyright 2020 ...

Page 8: ... electronic section LCC Respira is equipped with HEPA filter can be replaces with other filter in accordance to the use requirement able to sieve and filter micro particles and virus without stifling or suffocating the user The filter is replacable after certain length of usage ensuring safety in long term usage Clean air is pumped to the mask to ease breathing The electronic section employs an el...

Page 9: ...re 2 b and close the lid tightly Figure 2 c 3 Attach the CACP hose to mask Figure 2 d and to motor container Figure 2 c so as to connect all components Figure 3 a 4 Insert motor container Figure 2 c to the waist bag 5 LCC Respira 1 is ready to use Figure 3 b b b c d Figure 2 a Opening filter closure b Replacing filter c Ready to blower and motor d CPAP dan Masker ...

Page 10: ...e the battery after each use 3 Make sure all the components are perfectly connected 4 Turn the machine on using the on off button on elektronic container Figure 4 Figure 4 On off button and indicator lamp 5 As the motor is on air is filtered and blown 6 The filtered air is blown through CPAP hose to the mask 7 Check the sir condition within the mask whether it is stable 8 As the clean air stabiliz...

Page 11: ...ntainer tightly 5 Charge the battery until indicator lamp shown that the battery is full charged 6 The equipment is ready to reuse Figure 5 Battery reaplacement D Procedures to use LCC Respira 1 1 Make sure all components are complete and rightly installed 2 Put on the waist bag of electronic container on right and comfortable position Figure 6 3 Connect all parts 4 Put on the half mask to cover n...

Page 12: ...ustable 7 15 LPM Bacterial filter HEPA Blower material anti bacterial PLA Mask hose weight 150 gr Electronic box weight 480 gr Battery Rechargeable 3S Lithium Ion Battery Indikator alarm baterai Battery weight 320 gr Charging time Kurang dari 3 jam Operation tima low air flow 8 9 jam high airflow 5 7 jam Face protection Half Mask Filter blower elektronics ...