5.3 Calibration
Weight calibration is the correction of the difference between the
measured value displayed and the true weight (mass) of a sample, or the
reduction of the difference to a level within specified permissible error
limits.
Procedure:
1. Click
“
MENU
”
on the homepage, then click
“
CALIBRATION
”
to perform
the weight calibration operation.
2. Click the
“
Single point calibration
”
icon.
3. Wait until the value (
“
100.000
”
) of the desired adjustment weight
flashes and
“
please out on weight
”
displayed.
4. Put the calibration weight carefully into the center of the sample pan,
“
100.000
”
should stop flashing and
“
please waiting
”
displayed.
5. Wait until
“
please take away the weight
”
displayed, remove the
calibration weight.
6.
Wait until the page returns to the home page, the calibration is
completed.