2

stability in operation of the scale, be sure to ground the scale.

Avoid getting the scale wet as it is not water resistant. Any leakage of

liquid into the scale may damage the scale or cause an electric shock to

the user.

Use a power source (voltage, frequency, outlet type)

adapted to the specification of the scale. If excessive voltage

is used, the scale may overheat and be damaged or cause a fire.

Operate the scale on a stable, rigid and flat table.

Handle the scale carefully. It is a precision device,

subjecting it to impact may result in a malfunction.

If the device is not be used for a long period of time, the power

should be turned off and disconnect the power cable.

Do not disassemble, remodel or repair this product or accessories.

Do not weigh items larger than the scale's range, as this may damage

the load cell of the scale.

Summary of Contents for USS-DBS91

Page 1: ...Instruction Manual USS DBS91 Bench Scale ...

Page 2: ...eyboard 4 2 3 Display Panel 5 Unpacking 6 Before Using 7 4 1 Leveling the scale 7 4 2 Calibration 9 Operation 10 5 1 Basic Weighing 10 5 2 Unit Switching 11 5 3 Counting 11 Setting 13 6 1 Baud Rate 13 Maintenance 14 7 1 Precautions 14 7 2 Cleaning 15 7 3 Disposal 15 Troubleshooting 16 Technical Data 17 ...

Page 3: ...manual completely and follow the usage instructions before installation and operation as this will help you to make full use of the functions and performance of the USS DBS91 Bench Scale 1 1 Safety Precautions The U S Solid USS DBS91 Bench Scale qualifies as state of the art technology and complies with all recognized safety rules Improper use or handling however can result in damage and or injury...

Page 4: ...f excessive voltage is used the scale may overheat and be damaged or cause a fire Operate the scale on a stable rigid and flat table Handle the scale carefully It is a precision device subjecting it to impact may result in a malfunction If the device is not be used for a long period of time the power should be turned off and disconnect the power cable Do not disassemble remodel or repair this prod...

Page 5: ...3 Design and Function 2 1 Components 1 Weighing Pan 4 RS232 Interface 7 Power Inlet 2 Display Panel 5 Leveling Indicator 8 Power Switch 3 Keyboard 6 Leveling Feet ...

Page 6: ...on Enter the counting function Select the option in the setting of baud rate Tare or zero the scale Perform the baud rate set while powering on Switch the option in counting mode Switch the weighing unit Perform weight calibration ...

Page 7: ...ay Panel No Description 1 Indicates positive values Indicates negative values Stabilization indicator Disappearance indicates stabilization of scale 2 Weight value 3 Weight unit 4 Remaining capacity of the battery ...

Page 8: ...The following accessories are part of the standard equipment for your new USS DBS91 Bench Scale 1 Scale Main Body 1 Weighing Pan rectangular 1 Power Cable 1 Instruction Manual 1 Quality Certification Check the instrument for damage in transit Immediately inform the U S Solid customer service if you have any complaints or parts are missing ...

Page 9: ... 1 Leveling the scale Accurate horizontal positioning is necessary for repeatable measurements and exact results To compensate for small irregularities or tilts at this location the instrument needs to be leveled Adjust the leveling feet of the bench scale until the air bubble in the indicator is centered The level indicator is located under the cover towards the rear of the bench scale ...

Page 10: ...of air bubble Adjustment method up Turn both feet clockwise down Turn both feet counterclockwise Left Turn left foot counterclockwise right foot clockwise Right Turn left foot clockwise right foot counterclockwise ...

Page 11: ... Changes in ambient conditions 1 Press TARE to zero the scale in the weighing state 2 Long press the CAL button and CAL will be displayed 3 After that 20000 or 10000 will be flashing displayed then place the calibration weight on the pan 4 will be displayed then 20000 or 10000 will be displayed again 5 Remove the calibration weight and will again displayed 6 The calibration is completed and it wil...

Page 12: ...ing container on the weighing pan 2 Wait for the display to stabilize Then press TARE to tare and the displayed value will return to zero 3 Place the sample s to be weighed in the weighing container 4 After the display stabilizes read the value displayed Overweight Alarm When is displayed in the weighing state it indicates that the sample weight exceeds the maximum range of the scale The sample sh...

Page 13: ...unction that meets various industrial counting requirements especially the counting function for small components To ensure the accuracy of the counting function for small components it is necessary to ensure that the weight of all samples is consistent and that the weight of a single object is 4g 1 Press COU and COU will be displayed 2 After that 10 PCS will be flashing displayed 3 Press TARE to ...

Page 14: ... to be counted after the PCS again displayed and the corresponding number appears Exit counting function 1 Press COU and will be displayed 2 Then it will exit the counting function and return to the weighing interface ...

Page 15: ...t to change the transmission speed of the interface 1 Press TARE while powering on and the baud rate options will be displayed cyclically 2 When looping to the required baud rate press COU to confirm the selection 3 Then the baud rate set successfully ...

Page 16: ... to the weighing system do not let the object fall from the height onto the weighing pan 3 Do not expose the scale to high humidity or dusty conditions for an extended period of time 4 When the scale is not to be used for a long time cover it with to prevent dust from entering 5 When weighing powder and fine particles should be done with a suitable container to prevent dust and particles from fall...

Page 17: ...eaning agents solvents abrasive cleaning agents etc or organic solvents to clean the scale Clean the scale using a piece of lint free cloth which has been wet with a mild detergent soap only 7 3 Disposal Disposal of equipment and packaging must be carried out by the operator in accordance with relevant laws of the country or region in which the equipment is to be used ...

Page 18: ...ower transformer damaged Instrument faulty Problem Weight display is constantly changing or unstable Possible causes Test bench vibration Electromagnetic field interference Power supply instability Problem Displayed value and actual weight are not consistent Possible causes Not returned to zero before weighing Scale not leveled well ...

Page 19: ... Capacity 16 kg 21 kg 31 kg Readability 1g Repeatability 2 g Linearity Deviation 3 g Stabilization Time Within 3 seconds Pan Size 12 9 inches 315 230 mm Interface RS232 Power 110V 60Hz AC Dimensions 15 8 15 0 6 3 inches Net Weight 7 72 lbs Shipping Weight 8 82 lbs ...

Page 20: ...Contact Feel free to visit our website www ussolid com You can email us at service ussolid com You can call one of our friendly customer service representative at 1 800 209 4177 ...