U.S. Solid USS-DBS83 Instruction Manual Download Page 22

20

6.4 Percent Weighing

The U.S. Solid USS-DBS83 Analytical Balance has a built-in percent

weighing function.

1. Press

TARE

to zero the balance in the weighing state.

2. Press the

SET

button several times in succession until

PCT

is

displayed.

3. Press

TARE

and

PCT 100PCT

will be displayed.

4. Put the item to be used as a reference on the weighing pan and

confirm by pressing

TARE

.

5. "100.00 PCT" will be displayed. Then remove the reference sample

and put on the sample.

Summary of Contents for USS-DBS83

Page 1: ...Instruction Manual USS DBS83 Analytical Balance ...

Page 2: ...ing Components 9 3 4 Leveling the Balance 9 3 5 Connecting the Power 11 Settings 12 4 1 Display Speed 12 4 2 Stability 13 4 3 Restore Factory Defaults 14 Calibration 15 Weighing and Applications 16 6 1 Weighing 16 6 2 Unit of Weight 17 6 3 Counting Function 18 6 4 Percent Weighing 20 Maintenance 21 7 1 Precautions 21 7 2 Cleaning 22 7 3 Disposal 22 Troubleshooting 23 8 1 Troubleshooting 23 8 2 Err...

Page 3: ...se read the manual completely and follow the usage instructions before installation and operation as this will help you to make full use of the functions and performance of the USS DBS83 Analytical Balance 1 1 Safety Precautions The U S Solid USS DBS83 Analytical Balance qualifies as state of the art technology and complies with all recognized safety rules Improper use or handling however can resu...

Page 4: ... type adapted to the specification of the balance If excessive voltage is used the balance may overheat and be damaged or cause a fire Operate the balance on a stable rigid and flat table Handle the balance carefully It is a precision device subjecting it to impact may result in a malfunction If the device is not be used for a long period of time the power should be turned off and disconnect the p...

Page 5: ...3 Design and Function 2 1 Components 1 Weighing chamber 2 Weighing pan 3 Level bubble 4 Display panel 5 Keyboard 6 Glass door 7 Handle of glass door 8 Balance main body 9 Leveling foot ...

Page 6: ...4 1 RS 232 connector 2 Power inlet ...

Page 7: ... 2 2 Keyboard Key Description Turn the power on or off Enter settings and functions selection Tare Zero the balance Confirm selection Performs calibration procedure Switch between units Select mode item ...

Page 8: ...ription 1 Weight value 2 Indicates positive values 3 Indicates negative values 4 Stabilization indicator 5 Capacity utilization ratio 6 Weight unit 7 Animal weighing 8 Automatic calibration time mode 9 Display speed 10 Stability ...

Page 9: ...eteness The following accessories are part of the standard equipment for your new USS DBS83 Analytical Balance 1 Balance Main Body 1 Weighing Pan 1 Draft Shield 1 Power Cable 1 Calibration Weight 1 Instruction Manual 1 Quality Certification Check the instrument for damage in transit Immediately inform the U S Solid customer service if you have any complaints or parts are missing ...

Page 10: ... 30 C the relative humidity is around 50 and that non condensing conditions are met Avoid direct sunlight and ensure that there are not any excessive temperature fluctuations Ensure the balance is places a sufficient distance from heat sensitive materials in the vicinity of the instrument Avoid the effects of air currents from air conditioners ventilators open doors or windows Keep away from objec...

Page 11: ...r close 3 4 Leveling the Balance Accurate horizontal positioning necessary for repeatable measurements and exact results To compensate for small irregularities or tilts at this location the instrument needs to be leveled Adjust the leveling feet of the analytical balance until the air bubble in the indicator is centered The level indicator is located under the cover towards the rear of the analyti...

Page 12: ...of air bubble Adjustment method up Turn both feet clockwise down Turn both feet counterclockwise Left Turn left foot counterclockwise right foot clockwise Right Turn left foot clockwise right foot counterclockwise ...

Page 13: ...ource if it does not match Connecting a power cable a Insert the female end of the power cord into the power inlet located at the rear of the main unit b Plug the male end of the power cord into the outlet Note To obtain accurate results the analytical balance must be warmed up for at least one hour each time it is connected to an AC power source or after a power outage of more than 30 minutes Onl...

Page 14: ...ons Description Highest speed High speed Normal speed Low speed Depending on the environment in which the instrument is used carefully select the display speed to avoid instability that might be caused by excessive speed 1 Press the SET button several times in succession until the SPEED is displayed 2 Press TARE to enter the selection of display speed mode 3 Press UNIT to select required speed and...

Page 15: ...balance be adjusted appropriately in an environment that guarantees its normal operation Icons Description Lowest stability Low stability Normal stability High stability 1 Press the SET button several times in succession until the STAB is displayed 2 Press TARE to enter the selection of stability 3 Press UNIT to select required stability and confirm by pressing TARE ...

Page 16: ...nstable the stability of the balance can also be adjusted by restoring the factory settings 1 Press the SET button several times in succession until the DEFAULT is displayed 2 Press TARE to confirm and an audible signal will sound 3 Then it will back to the splash screen the factory defaults have been restored The setting is only applicable to situations where settings are confusing due to incorre...

Page 17: ... moving in the same room Changes in ambient conditions Prior to each use The balance must be fully warmed up for one hour to stabilize the weighing before calibration Procedure 1 Press TARE to zero the balance in the weighing state 2 Press the CAL button and the CAL 200 flashing displayed 3 Place the 200g calibration weight on the weighing pan 4 Then 200 0000g will be displayed External calibratio...

Page 18: ...dicates a stable state 4 Read the value displayed Weighing a sample that needs to be in a container 1 Open the glass door of the weighing chamber and place the weighing container on the weighing pan then close the glass door again 2 Wait for the display to stabilize The appearance of the stability mark indicates a stable state Then press TARE to tare and the displayed value will return to zero 3 O...

Page 19: ...an be used to meet the unit requirements in various usage situations 1 Press the UNIT button to select required unit 2 Press TARE to zero the displayed value and begin to weighing the sample To avoid the deviation of the weighing record caused by unit misreading please ensure that the selected unit is the one required ...

Page 20: ...the counting function for small components it is necessary to ensure that the weight of all samples is consistent and that the weight of a single object is 0 5mg 1 Press the SET button several times in succession until the COU is displayed 2 Press TARE to enter the counting function COU 5PCS will be displayed 3 Press UNIT to select the quantity and put the corresponding quantity of samples Press T...

Page 21: ... The operation can be started when the COU disappears and the corresponding number appears 5 After the test is finished press ON to select weight and press O T to confirm to exit and return to the weighing interface ...

Page 22: ... the balance in the weighing state 2 Press the SET button several times in succession until PCT is displayed 3 Press TARE and PCT 100PCT will be displayed 4 Put the item to be used as a reference on the weighing pan and confirm by pressing TARE 5 100 00 PCT will be displayed Then remove the reference sample and put on the sample ...

Page 23: ...height onto the weighing pan 3 Do not expose the balance to high humidity or dusty conditions for an extended period of time 4 When the balance is not to be used for a long time cover it with to prevent dust from entering 5 When weighing powder and fine particles should be done with a suitable container to prevent dust and particles from falling into the load cell below the weighing pan 6 Wear a g...

Page 24: ...e balance using a piece of lint free cloth which has been wet with a mild detergent soap only Removing the sliding glass doors for cleaning 1 Remove the pan and anti draft ring from the weighing chamber 2 Unscrew and remove the inside knob on the glass door 3 Slide the glass door out backwards 7 3 Disposal Disposal of equipment and packaging must be carried out by the operator in accordance with r...

Page 25: ...display is constantly changing or unstable Possible causes The sample pan is in contact with the draft shield or heating chamber Glass draft shield is not closed Test bench vibration Electromagnetic field interference Power supply instability Problem Displayed value and actual weight are not consistent Possible causes Not calibrated before weighing Not returned to zero before weighing Balance not ...

Page 26: ...24 8 2 Error Code Error code display Description Err Calibration operation error Err 1 Counting setting error Err 2 The weighing pan is not placed correctly Err 3 Sample weight exceeds the capacity ...

Page 27: ... 0 0001g 0 1mg Repeatability 0 0002g Linearity deviation 0 0003g Stabilization time Within 3 seconds Operating temperature 41 95 5 35 Pan size Φ80mm Interface RS232 Power 110V 60Hz AC Dimensions 11 8 7 9 11 6 inches 300 200 295 mm Net weight 13 23 lbs 6 kg Shipping weight 17 64 lbs 8 kg ...

Page 28: ...Contact Feel free to visit our website www ussolid com You can email us at service ussolid com You can call one of our friendly customer service representative at 1 800 209 4177 ...

Reviews: