10
VI. Using the Balance
Level adjustment
Once the balance is placed at its location, check the bubble level; if
bubble is not well-centered, turn the adjustable feet so the bubble
moves to the center and the balance is properly leveled.
Start
1. Connect the balance to the power supply and press ON key on the
panel. Wait 30 minutes after powering on for best performance.
2. After warming up, calibrate the balance.
3. After calibration, the balance will enter weighing mode
4. To turn the balance off, press the OFF key on the panel and the
display will shut off. If the balance will be unused for an extended
period of time, disconnect it from the power supply.
Calibration
1.
Press EXT CAL key; the value of the necessary weight will flash on
display
2.
Put the calibration weight on the pan. Wait until the value of the
calibration weight stops flashing and is fixed on the display (e.g.
100.0000 g, 200.0000g, etc., depending on the model)
3.
Remove the weight from the pan
4.
The balance will enter weighing mode and will display 0.0000g