II. Specific tions

Analytical balanc

es off

er fast and accurate measurements of 

mass, using strain gauge technology for fast, effici t mea-

surement.

Technical Parameters:

MODEL:

USS-DBS3-1

USS-DBS3-2

USS-DBS3-3

USS-DBS3-5

Max Capacity in 

grams

100

200

300

500

Accuracy Class

Class II

Min. Capacity

10 mg

Division

0.001g

Linearity

±0.002 g

Repeatability

±0.002 g

Operating 

Temperature

5° C- 35° C, with flu tuations of <1°C/Hour

41° F- 95° F with flu tuations of <1.8° F/ Hour

Relative Humidity

50%-85%

Power

AC 110V-240V, DC 9V

V. Using the Balance

Level adjustment

Once the balance is placed at its location, check the bubble level; if bubble is not well-centered, 

turn the adjustable feet so the bubble moves to the center and the balance is properly leveled.

Start

1. 

Connect the balance to the power supply and press ON key on the panel. Wait 30 minutes 

after powering on for best performance.

2. 

After warming up, calibrate the balance.

3. 

After calibration, the balance will enter weighing mode

4. 

To turn the balanc

e off

, press the OFF key on the panel and the displa

y will shut off

. If the 

balance will be unused for an extended period of time, disconnect it from the power supply.

Calibration

3.

Remove the weight from the pan

4.

The balance will enter weighing mode and will display 0.000g

.

Note: Each balance includes the necessary weight to complete external calibration

Tare function

1.

Put the receptacle on the weighing pan; its weight will be displayed

2.

Press TARE key and the reading on the display will automatically be set to 0.000g; tare is 

complete

3.

If the receptacle is removed from the pan, the corresponding weight, with negative value 

will be displayed.

4.

Press the TARE key again and the value on display will be set to zero again 

Weighing unit selection

Press UNIT key to select the desired weighing unit from g, ct, oz, and lb

Piece counting function

1. 

Press the PCS button

2. 

Press the CAL button repeatedly to cycle thr

ough diff

erent counts. This function can be used 

with counts of 10, 20, 50, 100, 150, 200, 250, or 500 pieces. 

3. 

Put the corresponding number of pieces on the pan, press the PCS button to confi m.

4. 

Remove the pieces from the pan and the balance is ready to be used in piece counting 

mode. 

5. To return to weighing mode, simply press the PCS button again.

1.

Press and hold down CAL key, when it shows ”CAL” on display, release the keys; then value 

of the necessary weight will flash on display.

2.

Put the calibration weight on the pan. Wait until the value of the calibration weight stops 

flashing and is fixed on the display (e.g. 100.0

0

0 g, 200.0

0

0g, etc., depending on the model).

Summary of Contents for USS-DBS

Page 1: ...e made available to all users of this equipment For best results and for maximum durability of the equipment carefully read and follow all instructions ANALYTICAL BALANCE USS DBS MODELS OPERATING MANUAL ANALYTICAL BALANCES ...

Page 2: ...ties and other similar canons Judicial surveys Pharmaceutical medicine preparations as well as analysis made in medical or pharmaceutical laboratories Determining the price or total amount in sale price or in pre Instructions on environmental protection At the end of its life cycle please do not dispose of this equip ment by throwing it in the usual garbage Instead hand it over at a collec tion po...

Page 3: ...ng the balance To prevent fi e and electrical issues avoid dusty or overly dry environments In case of electrical issue or fi e unplug the balance immediately During setup installation and use contact your wholesaler with any questions that arise This equipment is protected under the Warranties and Consumer Good Regulation 10 2003 Overhaul is not covered by the equipment warranty Changes made to t...

Page 4: ... be plugged into a grounded electrical outlet and the socket should be easily accessible to unplug in case of emergency Never unplug the balance by pulling on the wire do so from the base Never use the balance in a wedged in location such as a shelf Never use other objects such as pens pencils etc to press the buttons on the control panel use only your finger Never place anything on the pan heavie...

Page 5: ... balance will be unused for an extended period of time disconnect it from the power supply Calibration 3 Remove the weight from the pan 4 The balance will enter weighing mode and will display 0 000g Note Each balance includes the necessary weight to complete external calibration Tare function 1 Put the receptacle on the weighing pan its weight will be displayed 2 Press TARE key and the reading on ...

Page 6: ...y equipment c Direct sunlight or radiation 4 Do not use the balance anywhere exposed to explosive combustible or corrosive gases 5 Do not use the balance in areas with high humidity or high levels of dust 6 When moving from a cooler to a warmer place the accuracy and reliability of the scale will be influen ed by the moisture condensation inside the balance In order to eliminate this influen e it ...

Reviews: