PAGE 4 OF 4
0616 IH-2781
IMPORTANT! Wipe off stains on the spindle
sliding face with a dry cloth or a cloth
dampened with a small amount of alcohol.
Do not lubricate the spindle at this time.
• Wipe off stains on the dial cover or the digital
display surface with a soft, dry cloth or a cloth
dampened with a small amount of neutral
detergent. Be sure to use only a neutral detergent.
• To clean the indicator, use a soft cloth soaked in
a diluted neutral detergent. Do not use organic
solvent (thinner or benzene). It may deform or
damage the indicator.
• Deterioration of instrument performance is
caused mostly by harsh operating conditions. It is
recommended that the instrument be periodically
inspected, the frequency being determined by the
amount of use, environment and the way in which the
instrument is used.
MAINTENANCE
TROUBLESHOOTING
ERROR MESSAGE
OPERATING ISSUE
RECOMMENDATIONS
The battery is depleted.
Replace the battery.
Err C
Water drops may have formed on the detector
unit due to sudden thermal change, or the
detector unit may be contaminated due to
other causes.
If this message still appears after two hours of thermal
stabilization, the instrument needs to be repaired.
Contact Uline Customer Service at 1-800-295-5510.
E
An ABS data composition error occurred while
the spindle was moving at high speed.
Continue using the instrument, since this error does not
affect measured values.
If this error occurs while the spindle is stopped, it
could be due to sensor failure. In this case, contact
Uline Customer Service at 1-800-295-5510.
If the problem persists, or if the troubleshooting section does not resolve or describe your problem, contact
Uline Customer Service at 1-800-295-5510.