AutoGetWifi Intelligent Fiber Endface Microscope

Figure3-4

Figure3-5

3.2.4 Device software CCD Verification

3.2.5 About

Click the interface "Settings - CCD" to enter the CCD Verification interface, the purpose is to 
eliminate the inherent dirty spots on the CCD.

Click “Setting - About” on the interface to enter the about interface as follows:

Summary of Contents for AutoGet Wifi

Page 1: ...Please read this manual thoroughly before use Properly keep this manual with the equipment for future reference AutoGet Wifi Intelligent Fiber Endface Microscope...

Page 2: ...ver If there is no clear erroneous or ambiguous content founded in the manual please contact us 1 Keep the optics clean 2 Keep the monitor clean Prevent TFT LCD screen from the sharp cut or being crus...

Page 3: ...g part in accordance with this limited warranty No warranty will apply if 1 Device damage caused by wrong operation 2 Unconventional repair or modification 3 The causes of failure outside of the instr...

Page 4: ...tics and the basic operation of the equipment 3 AutoGetWifi Intelligent Fiber Endface Microscope Profile Hardware basic operation This chapter introduces software installation features interface and b...

Page 5: ...Replace the adapter 13 2 3 8 MT adapter use 14 3 1 1 Device software installation and update 14 3 1 2 PC software installation and update 14 3 2 1 Device software main interface 15 3 2 2 Device softw...

Page 6: ...1 PC software main interface 28 3 4 2 PC software settings 29 3 4 3 PC software FTP function 30 3 4 PC software interface 28 3 3 3 Image brightness setting 25 3 3 4 CCD Verification operation 26 3 3 5...

Page 7: ...the device The operating status is blue If the measurement passes it is green if the measurement fails it is red Power button Press and hold the Power Button to turn the power on Touch screen Display...

Page 8: ...gle adjustment button SD card slot Install SD card to store data and reports USB interface DC 5V Type C Battery status LED indicator Displays the battery power and the storage status during the chargi...

Page 9: ...croscope Figure1 4 Light Sensor The screen brightness can be adjusted according to the ambient light intensity Touch screen Display image and software operation interface Focus button Manual focus but...

Page 10: ...ment of the optical fiber end face at any time by data and report and store it on the SD card Projects 0 92um 0 52um 1 1 8 CMOS 5 8 72 ______ 480X800 3 2 LCD touch screen MT Parameters Common Interfac...

Page 11: ...tery and then pull it out Battery usage 1 Use a power supply to charge during the shutdown state of the device and it will take about 4 hours to charge 2 When the low light of the power indicator is f...

Page 12: ...was stored for a long period put it well in its half capacity and charge it again 3 months later Do not wrap it with conduct material to avoid the damage caused by the direct contact between the metal...

Page 13: ...AutoGetWifi Intelligent Fiber Endface Microscope 12 Figure2 4 Figure2 5 2 3 4 Install WiFi module 2 3 5 Measurement Install WiFi Module Press the measurement button to show real time image...

Page 14: ...switch 2 3 7 Replace the adapter Turn on the light switch to illuminate when the outside light is dark Replace the adapter 1 Loosen the lock nut of the adapter 2 Remove the adapter 3 Insert a new adap...

Page 15: ...ation and update 1 Copy the software that needs to be updated to the AutoGetWifidevice SD cardroot directory and then install the SD card to the AutoGetWifidevice card slot 2 Start the AutoGetWifidevi...

Page 16: ...system settings and measurement task settings ID ID number will be saved in data report file INC The ID number is automatically increased after the measurement AF On Auto focus switch Measurement res...

Page 17: ...into SM MM SF MF and are selected according to actual measurement requirements Scratch defect application After checking the measurement function of the spot or scratch of the current measurement area...

Page 18: ...task LCD brightness Users can manually adjust according to the needs of the actual use of the scene or you can check the Auto device to automatically adjust according to the brightness of the external...

Page 19: ...3 5 3 2 4 Device software CCD Verification 3 2 5 About Click the interface Settings CCD to enter the CCD Verification interface the purpose is to eliminate the inherent dirty spots on the CCD Click Se...

Page 20: ...o restart Software version information Displays the current software version number In the real time image display state press the device measurement button to measure After the measurement is complet...

Page 21: ...ope 20 Figure3 7 Figure3 8 1 Select a measurement task Before selecting the MT measurement task make sure that AF is ON and the Layer is ON The following is an operation example under task IEC2 0 PC M...

Page 22: ...o trigger image acquisition If the acquisition fails or there are other abnormalities please select the Redo or Cancel measurement button 4 Second acquisition image The MT interface rotation nut mark...

Page 23: ...on If the acquisition fails or there are other abnormalities please select the Redo or Cancel measurement button After 3 times of image acquisition the system automatically analyzes and displays the m...

Page 24: ...AutoGetWifi Intelligent Fiber Endface Microscope 23 Figure3 13 Figure3 14 Measurement failures can be selected by selecting Redo or Cancel 1 Unrecognized fiber 2 Failed to find the center...

Page 25: ...ition in strict accordance with the position of the rotating nut mark values 1 2 and 3 If the image is not collected in the required order the measurement will fail 2 Failure to detect all fibers will...

Page 26: ...ct SF and then adjust the image brightness The brightness of the SF image is recommended to be adjusted between 185 and 195 MF image brightness setting If the brightness of the MF image is not suitabl...

Page 27: ...ace After clicking the Verify button the prompt 1st image captured please rotate the fiber and capture again 2 Click OK rotate the ferrule and click the Verify button again prompting CCD Verifyfin ish...

Page 28: ...AutoGetWifi Intelligent Fiber Endface Microscope 27 Figure3 20 3 3 5 Layer 3 3 6 Image magnification and reduction Figure3 19 Layer on Layer off...

Page 29: ...1 3 4 PC software interface 3 3 7 File view The storage file format is data and report saved in the device SD card The file storage method is 1 Not 2 Data 3 Report 4 Data and report 5 pass data and re...

Page 30: ...The ID number is automatically increased after the measurement Connect Connect or disconnect wifi Measurement Trigger software measurement function Measurement results area PASS or FAIL Overlay Check...

Page 31: ...tting interface Data and report storage path SingleFiber data file name Default name MeasureDataSF MultipleFiber data file name Default name MeasureDataMF About interface Language selection Support Ch...

Page 32: ...on the main interface of the PC software 4 Open SD card and save the data file in the Measure Result folder If you have any further questions please contact Dimension Technology Co Ltd Tel 86 755 2648...