SECTION C – SERVICE INFORMATION
13
GENERAL MAINTENANCE, INSPECTIONS & STORAGE
The expected service life of the
APT-5
is at least 7 years.
Minimum ongoing maintenance should prevent unnecessary repairs.
The rugged design of the
APT-5
and the use of selected, modern materials ensure minimal
requirements for care and maintenance.
NOTE: Improper handling or neglect in the care of the
APT-5
may reduce or cancel the
coverage of the manufacturer’s warranty.
13.1 Regular care
Monthly checks
• Inspect Power supply cables and plug for visible damage.
• Check power-input connector for visible damage or insecure fastening.
• On a regular basis check that all screws and components are fastened tightly.
• Ensure that the anti-slip pads under the base are always kept clean.
Annual checks – Checks after repair
These annual checks shall be performed by an authorised technician.
• Ensure full functioning of Operator Panel (Fig. 3, page 10)
• Smooth rotation of Crank Arms (Fig.1.4 , page 8)
• Firm locking of Angle Securing Knob (Fig. 1.3 page 8)
• Ensure that the Input Socket (Fig. 1.5 page 8) is not damaged.
• Inspect the power supply for the following parameters:
Parameter
Test Conditions
Min.
Typ.
Max.
Unit
Safety Ground Leakage Current Io=Full Load, Vin=240 VAC/60Hz
0.1
mA
Earth Impedance
25 A
40 A
22
27
m
Ω
Isolation Resistance
Test Voltage=500
VDC
50
m
Ω
CAUTION: If any damage is detected – do not use the
APT-5
. Please contact your authorized
dealer. Only authorized personnel may carry out repairs. The power supply is not to be
repaired.
13.2 Cleaning instructions
• Disconnect Power Supply and wipe dry with clean cloth.
• Take care not to allow water to enter the unit. Keep cables and electric components
away from water and humidity.
CAUTION:
For safe disconnection of the
APT-5
always take the mains plug out of the wall
socket before removing the connector from the
APT-5
power socket.