Operation Temperature

50°F ~ 104°F (10°C ~ 40°C)

Operation Humidity

10 ~ 90% RH

Hematocrit

10 ~ 65%

Test Sample

Capillary Whole Blood

(fingertip, palm, and forearm)

Sample Volume

0.7 µL

Measuring Unit

mg/dL

Measuring Range

20 ~ 600 mg/dL

Test Time

5 seconds countdown

Memory Capacity

500 most recent results

Average

7/14/30 days average

External Output

USB interface

Power supply

2  AAA batteries

Battery Life

Approximately 1000 tests

Dimension

106L x 66W x 20H mm

Weight

70 g without batteries

Specifications

Note: Please refer to the Test Strip Insert for 
the performance of system accuracy and 
precision.

The Meter meets the electromagnetic 

compatibility requirements as per IEC 

60601-1-2:2014 and EN 60601-1-2:2015.

 

Caution: Do not use the meter close to the  
electronic equipment which is a source of 
strong electromagnetic radiation, which may 
interfere with proper operation of meter.

34

Summary of Contents for HT100-C

Page 1: ...s experiencing a hyperglycemic hyperosmolar state with or without ketosis For In vitro diagnostic only For over the counter use Alternative site testing should be performed only during steady state wh...

Page 2: ...cose Test Strip Tyson Bio HT100 C Blood Glucose Test Strip Setting Up Tyson Bio HT100 C Blood Glucose Meter Installing Replacing Batteries Date Time Setup Reminder Alarms Setup HI LO Alarms Setup Impo...

Page 3: ...trol Solution Results Memory Features View Day Average Results Exiting the Memory Mode Transfer Test Results to a Computer Voice Guidance Bluetooth Guidance Cleaning and Disinfection Procedure Care an...

Page 4: ...the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips forearm or palm Alterative site testing should be performed only during steady state when gluc...

Page 5: ...sinfecting Solution Clorox Bleach Germicidal Wipes Clorox Professional Products Company EPA Reg No 67619 12 Reference FDA Public Health Notification Use of Fingerstick Devices on More than One Person...

Page 6: ...he complete kit contains HT100 C Blood Glucose Meter HT100 C Blood Glucose Test Strip 10 strips HT100 C Control Solution Level 2 Adjustable Automatic Lancing Device 10 Sterile Lancets User Manual Quic...

Page 7: ...o enter buzzer setup To adjust numbers to toggle settings To browse test results and averages in memory 5 Button Press and hold to enter the year date time setting To browse Control Solution test resu...

Page 8: ...N 7 Meal Indicator Icon Pre meal and post meal 8 14 day Average 14 day average result 9 Time Time display 10 Blood Drop Icon Flashes when the meter is ready to accept your blood sample to test 11 Warn...

Page 9: ...ucose Meter Use each strip ONCE ONLY A test strip consists of the following parts 1 Target Area Where a blood drop or control solution to be applied 2 Confirmation Window To confirm sufficient blood s...

Page 10: ...mg dL Level 2 mg dL Level 3 mg dL Tyson Bio HT100 C Blood Glucose Test Strip Test Strip Vial Label Test Strip Vial Blood Glucose Test Strips HT100 C Control Solution Range Store at 39 86 F 4 30 C 10...

Page 11: ...and flash Press or to adjust and press to confirm Fig 7 7 The numbers of hour will appear and flash Press or to adjust and press to confirm Fig 8 Installing Replacing Batteries 1 Slide and remove the...

Page 12: ...ext reminder alarm setting If ON is chosen Fig 13 and pressed the number of hour will start flashing Fig 14 Press or to adjust and press to confirm 5 The numbers of minutes will flash Fig 15 Press or...

Page 13: ...ion and HI Alarm on you will hear please set hyperglycemic value 4 Continue to the LO alarm setting Fig 19 repeat Step 2 and 3 Fig 20 21 to confirm or to change LO alarm setting The LO alarm value ran...

Page 14: ...ution if the package is damaged Store your test strips and the Control Solutions between 39 F 86 F 4 C 30 C Temperature beyond this range can damage test strips and lead to inaccurate results Any chan...

Page 15: ...voice function on you will hear Please apply blood sample You can press or to select the meal indicator for marking the test as Fig 26 Pre Meal Test with the voice function on you will hear Pre meal...

Page 16: ...o unload the lancing device 1 After sampling carefully remove the cap containing the used lancet 2 Push the exposed tip of the lancet into its Protective Cap Fig 33 Slide the Lancet Ejector forward an...

Page 17: ...the test strip Fig 34 Discard the first drop of blood sample to prevent contamination 2 Hold the tip of the test strip touching the blood drop until the meter beeps 3 The meter will start to countdow...

Page 18: ...Fig 39 appears on the display it indicates the reading is higher than 600 mg dL If LO Fig 40 appears on the display it indicates the reading is lower than 20 mg dL We suggested the testing procedure...

Page 19: ...poglycemia low blood glucose we recommend that you test on your fingertips There are limitations for doing AST Please consult your healthcare professional before you do AST AST measurements should not...

Page 20: ...into the test port The meter will be automatically turned ON Fig 41 3 A beep will sound and the system checking will appear followed by a flashing icon Fig 42 with the voice function on you will hear...

Page 21: ...p Hold until the meter beeps The meter will now start to count down and the Control Solution test result will appear on the LCD screen Fig 45 with the voice function on you will hear Control solution...

Page 22: ...roll forwards and backwards 4 Press to exit and view the Control Solution test result Fig 48 5 When the meter displays the last stored test result press so the meter will display day average results F...

Page 23: ...displays 7 days with no meal status assigned press button and the meter will change to view results in memory View Day Average Results Fig 50 Fig 51 Fig 52 Fig 53 Fig 54 Fig 55 Fig 56 Fig 57 Note 1 Th...

Page 24: ...ter Obtain the required software and cable to be ordered separately For order information please call Customer Service or visit website of www tysonbio com 1 Follow the instructions provided with soft...

Page 25: ...nto setting mode 2 Press or to select ON Fig 60 or OFF Fig 61 A beep will sound when voice is set ON 3 Press to confirm 4 If the voice is set on press or to select voice volume The numbers of volume w...

Page 26: ...enter Bluetooth mode will appear and flash on the meter when meter is not connected with a mobile device Fig 65 2 Turn on BLE tool APP with Blood Glucose Test and Bluetooth on your mobile device 3 Cli...

Page 27: ...meter is NOT connected with a mobile device 6 Turn on BLE tool APP with Blood Glucose Test and Bluetooth on your mobile device Note Please skip Step 1 4 if Bluetooth feature is already ON 7 Click Con...

Page 28: ...ed with a mobile device 6 After testing your blood glucose the test result is shown Fig 72 Bluetooth icon is still blinking Note Please skip Step 1 4 if Bluetooth feature is already ON 7 Turn on BLE t...

Page 29: ...o 67619 12 Signs of deterioration Performance The control solution measurements are out of range which is listed on the test strip label Physical Polymer crazing thin silver streaks appear cracking sw...

Page 30: ...allow water or cleaning fluids into the meter test port and data port Disinfection Cleaning solution and disinfecting solution Clorox Bleach Germicidal Wipes Use a new disinfection wipe The meter is...

Page 31: ...eter test strips and the Control Solution to extreme conditions such as high humidity heat freezing cold or dust DO NOT STORE IN DIRECT SUNLIGHT OR AREAS WITH HIGH HUMIDITY AND OR DUST It is advised t...

Page 32: ...setting Display Description Action Test result is HIGHER than 600 mg dL Repeat the test by using a new test strip If the same message displays again the result is confirmed to be higher than the mete...

Page 33: ...rm a test 2 Use control solution to check the meter and test strips 1 Follow the instructions and try again with a new test strip Please contact customer service if the problem persists 2 Follow the i...

Page 34: ...power is depleted for further usage with the voice function on you will hear Battery error Replace with 2 AAA batteries immediately An error message indicates a problem with the test strip or meter wi...

Page 35: ...r 1 minute Reinstall batteries to see if meter is working properly Please contact customer service if the problem persists An error message indicates a problem with the test strip with the voice funct...

Page 36: ...than 2 minutes before testing Review the instructions and re insert a test strip correctly LCD display on the meter is blank when trying to perform a test Please contact customer services for assistan...

Page 37: ...USB interface Power supply 2 AAA batteries Battery Life Approximately 1000 tests Dimension 106L x 66W x 20H mm Weight 70 g without batteries Specifications Note Please refer to the Test Strip Insert f...

Page 38: ...use Serial Number Manufacturer In vitro diagnostic medical device Consult instructions for use Sufficient for Catalogue number Control Caution Transport and store at temperature 4 F 122 F 20 C 50 C le...

Page 39: ...offers 3 years guarantee on this product Our company shall repair or replace any Tyson Bio HT100 C Blood Glucose Monitoring System found defective with a new one This warranty does not apply to the p...

Page 40: ...uency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in...

Page 41: ...ice must accept any interference received including interference that may cause undesired operation of the device FCC RF Radiation Exposure Statement 1 This Transmitter must not be co located or opera...

Page 42: ...Lianfa Medical Plastic Products Co Ltd No 1 Shuangshan Sanjian Road 250200 Zhangqiu City Jinan Shandong PEOPLE S REPUBLIC OF CHINA SteriLance Medical Suzhou Inc No 68 Litanghe Road Xiangcheng Suzhou...