kANGAROO
™ 924
10
Enteral Feeding Pump
Load Set
(LoAd SEt)
Pump set has been improperly placed onto pump.
– Check to make sure only KANGAROO 924 Pump Set has been placed onto the pump. Check to make sure OPTIX connector
on pump set has been properly positioned into the OPTIX lock on pump.
High Ambient Light Condition
– If dots appear with the LoAd SEt message on the display, this indicates a high ambient light condition. Remove from
bright light area until error condition clears.
Feeding solution or other foreign substances have accumulated in the OPTIX lock and blocked or obscurred the
optical set detection mechanism.
– Follow the instructions in Section IV: Cleaning under “Directions for Cleaning OPTIX Lock” to remove the material
blocking proper detection of the feeding set.
System Error
Alarm (SYS Err)
Rotor is not turning while pump is running.
– Return pump for technical service.
Rotor is turning but is not detected by rotor sensors.
– Return pump for technical service.
VTBD Delivered Alarm
(Vtbd dEL)
Preset VTBD has been delivered to patient.
No Audible Alarm
Battery is fully discharged.
– Charge battery for a minimum of 12 hours.
Alarm circuitry failure.
– Return pump for technical service.
The KANGAROO 924 Feeding Pump is designed to minimize the effects of uncontrolled electromagnetic interference and other types
of interference. When using the KANGAROO 924 Feeding Pump, avoid the use of equipment that causes erratic operation or degradation
in performance.
Biotech Mode
To enter Biotech mode:
1. Turn pump off.
2. Press and hold the
ON
key for 6-7 seconds.
3. Pump alternately displays:
biotECH
for 5 seconds. Pump then displays
SEt LAng
for 5 seconds.
To select the language press or buttons or rotate knob to cycle through the list. When the desired language
is displayed, press the
ON
key to set.
4. Pump now displays:
bEEP LVL
and (volume of speaker)
Select desired volume (1 to 10) by rotating knob or pressing or buttons.
5. Pump returns to
HOLD
mode.