background image

kANGAROO

 924

Enteral Feeding Pump

Table of Contents

SECTION I 

General Information .................................................................................................................

SECTION II 

Operating Procedures ...............................................................................................................

 

 

Battery ....................................................................................................................................

 

 

Beginning Administration .........................................................................................................

SECTION III 

Alarms .....................................................................................................................................

SECTION IV 

Cleaning ...................................................................................................................................

 

 

Important User Information ......................................................................................................

 

 

General Cleaning Directions ......................................................................................................

 

 

Directions for Cleaning Pump Housing .......................................................................................

 

 

Directions for Cleaning Pump Power Cord ...................................................................................

 

 

Directions for Cleaning Drop Detector ........................................................................................

 

 

Directions for Cleaning OPTIX Lock .............................................................................................

 

 

Directions for Cleaning Rotor Assembly ......................................................................................

 

 

Preventive Maintenance ...........................................................................................................

SECTION V 

Troubleshooting Guide ..............................................................................................................

 

 

Biotech Mode ...........................................................................................................................

SECTION VI 

Factory Service .........................................................................................................................

SECTION VII 

Specifications ...........................................................................................................................

 

 

Symbols ...................................................................................................................................

SECTION VIII 

Warranty ..................................................................................................................................

English

Status: Review Routing On: CR10000186 Doc#: AW10001678

11/10/2006 Artwork#s: WN 182493-01 Job#: 14761

Summary of Contents for KANGAROO 924

Page 1: ...Directions for Cleaning Pump Housing Directions for Cleaning Pump Power Cord Directions for Cleaning Drop Detector Directions for Cleaning OPTIX Lock Directions for Cleaning Rotor Assembly Preventive Maintenance SECTION V Troubleshooting Guide Biotech Mode SECTION VI Factory Service SECTION VII Specifications Symbols SECTION VIII Warranty English Status Review Routing On CR10000186 Doc AW10001678 ...

Page 2: ...an 5 minutes Pump set is improperly loaded Self diagnosed problem which requires technical servicing Pump motor malfunction Ambient light error Automatic systems check after pump has been turned ON Large LED display Low occlusion pressure Pump will not overcome back pressure greater than approximately 12 psi 82 7 kPa Integral pole clamp Use of OPTIX pump sets to prevent free flow from inadvertent ...

Page 3: ...ip Chamber Guide 10 Pole Clamp Knob including drop sensor 11 AC Light 3 Tubing Guide 12 Vol CLR Light 4 Lower Drip Chamber Guide 13 Adjustment Knob 5 OPTIX Lock 14 Sensor Lens 6 Rotor Assembly 15 Tubing Latch 7 Touch Panel 8 LED Display and Pump Running Indicator Lights Status Review Routing On CR10000186 Doc AW10001678 11 10 2006 Artwork s WN 182493 01 Job 14761 ...

Page 4: ...lume delivered when pressed afterVOL has been pressed TheVOL CLR light will go on when volume delivered has been reset to zero Volume delivered cannot be cleared until the presetVTBD has been delivered or cleared VTBD VolumeTo Be Delivered When pressed allows user to preset a volume to be delivered by pressing or or rotating adjustment knob Holding for 5 seconds will clearVTBD setting and volume V...

Page 5: ...attery pack only with the same type and rating of battery Beginning Administration Below are general instructions for inserting allTyco Healthcare Group LP Pump Sets into the KANGAROO 924 Enteral Feeding Pump 1 Press ON 2 Wait while pump performs system check Displays for a moment 3 Check VOL CLR light or press VOL to insure that volume delivered is clear if desired 4 Rate of delivery may be chang...

Page 6: ...amp 15 Press START HOLD 16 Proceed with feeding 17 Dispose of pump sets Not to exceed 24 hours of use Used pump sets shall be handled as follows To prevent employee exposure risk the materials must be handled in accordance with bloodborne pathogen standards before being placed in an approved disposal container Disposal must be performed in accordance with current medical practices or local regulat...

Page 7: ...the system the occlusion has occurred Correct occlusion condition Press START HOLD switch Continue with feeding C The drop sensors are blocked and will not be able to detect drops as they come through the drip chamber To correct this condition check drop sensors on inside of upper drip chamber guide and clean off any material that may be blocking sensors D Tubing is pinched or kinked Low Battery L...

Page 8: ...d on their knowledge of the environment in which the pump is used Cleaning may become necessary between intervals due to unforeseen soiling Only personnel trained in the cleaning of medical devices shall perform cleanings Important User Information CAUTION DO NOT IMMERSE PUMP OR POWER CORD IN WATER OR OTHER CLEANING SOLUTION CLEAN USING A DAMP NOT WET CLOTH OR SPONGE FAILURE TO FOLLOW THE CLEANING...

Page 9: ...e power cord with a cloth dampened with isopropyl alcohol CAUTION Avoid exposing power cord to excess moisture as this can lead to an electrical shock or fire hazard Directions For Cleaning Drop Detector Refer to General Cleaning Directions before starting Clean the drop sensors in the upper drip chamber guide with a cotton swab dampened with isopropyl alcohol Be sure sensor areas are clear Direct...

Page 10: ...g has been in use for 24 hours replace Feeding tube or feeding set tubing is occluded Locate point of occlusion and correct Note occlusion may occur upstream or downstream from pump Feeding Container Feeding Pump Set control clamp closed FeedingTube patency of feeding tube should be checked Drip chamber is improperly placed in pump Check to make sure drip chamber is properly locked in position in ...

Page 11: ...y rotor sensors Return pump for technical service VTBD Delivered Alarm Vtbd dEL Preset VTBD has been delivered to patient No Audible Alarm Battery is fully discharged Charge battery for a minimum of 12 hours Alarm circuitry failure Return pump for technical service The KANGAROO 924 Feeding Pump is designed to minimize the effects of uncontrolled electromagnetic interference and other types of inte...

Page 12: ...ad Watertown NY 13601 Pointe Claire Qc Ballymoney BT53 7AP Northern Ireland H9R 5H8 B To place an order for repair or if technical assistance is required call customer service Technical information will be made available on request to allow the users appropriately qualified technical personnel to repair the parts of the equipment designated by the manufacturer as repairable Refer to the KANGAROO 9...

Page 13: ...lychargeswhenpumpispluggedintowalloutlet Rechargetimeforfullydischargedbatteryisapproximately12hours A new battery when fully charged will operate for approximately 16 hours at 125 milliliters per hour Dimensions Approximately 7 1 4 High x 5 1 2 Wide x 4 1 4 Deep Weight Approximately 3 625 lbs Case Material ABS PC Plastic FIRE RESISTANT Power Requirements 120V 60 Hz 1 Amp North America 240V 50Hz 0...

Page 14: ... replacement within the terms of the warranty The defective pump should be promptly returned properly packaged postage prepaid Loss or damage in return shipment toT H G L P shall be at Purchaser s risk This warranty shall not apply to andT H G L P shall not be responsible for any loss arising in connection with the purchase or use of any such pump which has been repaired by anyone other than an au...

Page 15: ...761 ARTWORK NO 182493 PACKAGETYPE DOM CD manual SIZE 8 5 X 11 PRINTED 1 SIDE INK COLOR Black BAR CODE INK COLOR N A Status Review Routing On CR10000186 Doc AW10001678 11 10 2006 Artwork s WN 182493 01 Job 14761 ...

Reviews: